Abstract
Abundant studies have been published evaluating different parameters of reverse-phase liquid chromatography (LC) and supercritical fluid chromatography (SFC), both coupled to electrospray (ESI)/mass spectrometry (MS) for pesticide residue analysis. However, there is a lack of a comprehensive comparative study that facilitates deep knowledge about the benefits of using each technique. In the present study, the same mass spectrometer was used coupled to both liquid and supercritical fluid chromatographies with a multiresidue method of 215 compounds, for the analysis of pesticide residues in food samples. Through the injection of the spiked extracts, separate experiments were conducted. A study of the optimum ion source temperature using the different chromatography modes was performed. The results were evaluated in terms of sensitivity with tomato, leek, onion, and orange as representative fruit and vegetable matrices. The compounds which reported the highest area values in each chromatography were evaluated through their substance groups and polarity values. The impact of matrix effects obtained in tomato matrix was similar for both cases; however, SFC clearly showed better results in analyzing matrices with a higher number of natural co-extracted compounds. This can be explained by the combination of two effects: (i) chromatography separation and (ii) ion source efficiency. The chromatographic elution presented different profiles of matrix components, which had diverse impact on the coelution with the analytes, being more beneficial when SFC was used in the matrices studied. The data showed that the best results obtained in SFC are also related to a higher ionization efficiency even when the ESI emitter tip was not optimized for SFC flow. In the present study a comprehensive evaluation of the benefits and drawbacks of these chromatography modes for routine pesticide residue analysis related to target compounds/commodities is provided.
Graphical abstract
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Since the introduction of modern mass spectrometer devices, reverse-phase liquid chromatography (LC) has been used as the prevailing technique for the analysis of relatively polar and thermolabile pesticides [1, 2]. The system robustness and the broad scope covered made it one of the preferred approaches to work coupled to mass spectrometry in routine laboratories. On the other hand, supercritical fluid chromatography emerged in 1962 as a technique with much potential; however, the drawbacks and limitations were high due to the lack of devices necessary to perform an efficient analysis. During the consecutive decades, many milestones helped improve SFC performance [3, 4]. Numerous studies highlighted the kinetic performance of SFC like the enhanced linear velocity compared to liquid chromatography. This fact can be easily observed through the van Deemter diagram, where the particular properties of the supercritical fluid provide an improved diffusion coefficient and solvent strength compared to LC [5]. In SFC, the elution profile is directly related to the density of the mobile phase; for this reason, the development of an efficient back pressure regulator (BPR) device that allows kee** the outlet pressure at the desired value is one of the most essential achievements in the SFC evolution [6]. The different couplings of the BPR in the SFC system have been studied since its implementation [4, 7]. Apart from that, the use of a co-solvent for carbon dioxide during the elution ushered a wide range of possibilities [8, 9]. The use of modifiers can favor the solubility of some compounds, improving chromatographic quality. Furthermore, the use of additives in the modifier enhances the interactions with the column stationary phase and the pH stability of the mobile phase [9]. Apart from the high number of possible combinations for the mobile phase in SFC, the interactions with the stationary phases usually present differences compared to LC. West et al. analyzed 109 compounds using 31 distinct stationary phases with the aim of facilitating the column selection in SFC [10]. The numerous advances and achievements in the last decades provided SFC with the necessary quality to be an alternative to the conventional liquid chromatography.
Supercritical fluid chromatography coupled to mass spectrometry has been used for the analysis of compounds in different fields, including environmental and food analysis. The online extraction coupled with SFC-MS/MS provides a fast and reliable determination of natural products like carotenoids and apocarotenoids in food and biological matrices [11]. This type of extraction is also hyphenated to SFC-MS/MS for the analysis of polycyclic aromatic hydrocarbons (PAHs) in soil matrix [12]. SFC coupled to high-resolution mass spectrometry (HRMS) provides a robust analytical method for the simultaneous quantification of persistent and mobile organic substances [13]. Moreover, this type of hyphenation of HRMS with SFC is often used in the metabolomic field where the matrix effect reduction seems to improve the analysis in complex matrices like urine or plasma [14]. However, one of the most important applications of supercritical fluid chromatography is the analysis and quantification of stereoisomers in chiral compounds [15]. In the last decades, this enantiomer separation with SFC has remarkably increased in the pharmaceutical industry [16]. Furthermore, preparative LC is being replaced by preparative SFC due to the advantages of using carbon dioxide as the mobile phase: high flow rate, short equilibration time, no dilution effect, and lower solvent consumption [17].
Regarding pesticides, the interest in using SFC to detect and quantify these residues has increased in recent years [18]. Some evaluations of multiresidue methods have been performed using tandem mass spectrometry [19, 20] and HRMS [21, 22]. As pesticides are often chiral compounds, enantioseparation of pesticides by SFC-MS/MS was previously researched [23,24,27]. As can be observed in Figure 1, the interface temperatures tested were 350 °C, 200 °C, and 125 °C. Very similar results were obtained with pesticides in a pure solvent at the three temperatures studied: all the compounds were identified in the 95–99% range in both chromatographic methodologies. We had almost an identical situation with the tomato matrix, when 92–95% of the pesticides were identified at the different temperatures using both chromatographies. Regarding the onion matrix, there were no vast differences between techniques (80–89% range) being the results slightly better using SFC at 200 °C. In leek matrix, using LC, 80% of the compounds were identified at 200 °C and 350 °C; however, 69% of the compounds were identified at a lower temperature of 125 °C. In SFC, the percentages of identified compound in leek were between 83–85% in all the temperatures. A similar situation was found in the orange matrix analyzed by LC, with percentages of identified compounds at 200 °C and 350 °C of 78% and 79%, respectively. Analogous to what happened with the onion matrix, in orange, there was a decrease of identified compounds at 125 °C, where the percentage of identified compounds was 67%. Again, the interface temperatures using SFC did not have a strong impact on the results. The percentages of identified compounds in orange using this technic were in the 86–89% range in all the temperatures tested. Considering LC, the highest temperature provided better results in all the matrices, obtaining lower areas as the temperature decreased. This data confirms that high ESI interface temperature favors ionization and provides a general increase of sensitivity in the LC pesticide multiresidue methods. This increase of sensitivity of analytes at higher temperatures using LC coupled to ESI sources has been described many times before [28, 29]. Faster drying of the droplets allows reaching the Rayleigh limit faster, and therefore produces more progeny droplets at the same time. In the case of ESI droplets with high organic solvent content, as is the case of SFC and late eluting compounds in LC, a higher drying gas temperature does not increase this drying rate as much as it does for water-rich droplets [30]. Regarding SFC, although there were a higher number of identified compounds at 350 °C, a similar percentage was obtained at the lowest temperature tested (125 °C). Unlike LC, no significant sensitivity differences were observed between the three temperatures tested.
Focusing on SFC area values, the differences between areas using the highest and lowest temperatures were lower than 30% in nearly all the compounds studied. Therefore, SFC did not show a very pronounced preference for higher or lower ESI interface temperature. This can bring additional benefits by analyzing some complex/thermolabile compounds at low temperatures, as previously demonstrated with captan and folpet [31]. Focusing on figure 2, we can observe the non-identified compounds per technique and matrix at the temperature test of 200 °C. Considering the number of pesticides not identified in each matrix, in LC we have an increase of approximately 30% per matrix compared to SFC, except in the orange matrix, where the non-identified compounds were almost twice in LC. Regarding solvent results, avermectin b1a was the only pesticide undetected in both chromatographies. The other two pesticides non-identified in SFC were cyazofamid and malathion (both presented isobaric interferences in the second transition). Considering LC, in addition to avermectin b1a, 5 pesticides were not identified: Bendiocarb and haloxyfop showed interferences in the second transition, iprodione presented low sensitivity and it decreased as the interface temperature was reduced; EPN and methamidophos also presented low sensitivity and they did not fulfill the ion ratio criteria at that low concentration level.
It is important to note that the spray emitter electrode used in the electrospray source is specifically designed for LC systems. Additional improvements in sensitivity can be expected by using a smaller diameter emitter electrode [32, 33]. Sampling efficiency in ESI using SFC is better due to the low flow reaching the source. In LC, the solvent flow is 300 μL/min; on the other hand, 75% of compounds in SFC elutes with a solvent flow below 140 μL/min. However, this mobile phase flow modification is not a dramatic reduction compared to LC typical flows. Moreover, there is a difference in the composition that arrives at the ESI source of the mobile phase depending on the type of chromatography. In the case of LC, the mobile phase starts at 100% water and increases to 100% of methanol during the elution, while in the case of SFC the elution in all cases is nearly 100% MeOH. Additionally, considering the gradient used in LC during the experiment, approximately 60% of the compounds eluted with 100% of methanol in the mobile phase and thus the inconveniences related to the presence of water during the ionization affected to a reduced percentage of pesticides in the multiresidue method. These factors could explain that even though the results were better using SFC, not huge sensitivity differences could be observed between both methodologies.
Chemical groups and polarity
Another comparative study was performed in pure solvent to determine if there were physicochemical parameters that affected the sensitivity of compounds in each type of chromatography. A vial with a mix of pesticides at the concentration of 5 μg/L in solvent was injected in both techniques. The areas of the pesticides were compared, and those with a difference higher than 50% between methodologies were considered as more sensitive in the technic with the higher value and they were evaluated in terms of polarity and chemical group. Regarding the compounds in LC, 70 pesticides presented an area value higher than 50% compared to the same compound’s SFC values. From these compounds, there were a wide variety of chemical groups. However, organophosphates were the predominant group, with 19 pesticides (27%): azinphos-methyl, chlorpyirifos, demeton-s-methyl, diazinon, ethion, ethoprophos, fenthion, malathion, methidathion, phenthoate, phosmet, phoxim, pirimiphos-methyl, profenofos, prothiophos, and quinalphos. The following most prevalent groups were carbamates with 10 cases (carbaryl, fenobucarb, isoprocarb, methiocarb, methomyl, metolcarb, pirimicarb, promecarb, propoxur, XMC), urea derivatives herbicides with 7 pesticides (chlorbromuron, chlorotoluron, fenuron, isoproturon, linuron, metobromuron, and monolinuron), and triazines with 6 cases (atrazine, prometryn, propazine, simazine, terbuthylazine, terbutryn). The remaining pesticides had 3 or fewer matches per substance group. On the other hand, using SFC, 77 pesticides showed differences in their chromatographic areas higher than 50% compared to LC. In this case, triazoles were the most representative chemical group, with 17 compounds (22%): bromuconazole, cyproconazole, difenoconazole, diniconazole, epoxiconazole, fenbuconazole, fluquinconazole, flusilazole, flutriafol, hexaconazole, metconazole, myclobutanil, paclobutrazol, propiconazole, tebuconazole, tetraconazole, and triticonazole. Triazoles were followed by 11 organophosphates (demeton-s-methyl sulfone, EPN, fenamiphos, fenamiphos sulfone, fenamiphos sulfoxide, fenthion sulfone, fenthion sulfoxide, methamidophos, omethoate, and pyridaphention), 7 benzoylureas (chlorfluazuron, diflubenzuron, flufenoxuron, hexaflumuron, lufenuron, novaluron, and triflumuron), and 5 neonicotinoids (acetamiprid, imidacloprid, nitenpyram, thiacloprid, and thiamethoxam). The remaining chemical groups were not repeated more than 3 times.
It is already known that the compound elution in SFC does not follow the same trend as it does with LC [20]. While in reverse-phase LC the compounds elute in decreasing order of polarity, in SFC there are other factors involved in retention mechanisms like fluid density and modifier interactions [34], as supercritical CO2 is a non-polar mobile phase (a modifier gradient of MeOH gradually changes this situation). The pesticides that showed a higher sensitivity for each technique (70 pesticides in LC and 77 pesticides in the case of SFC) were evaluated using the octanol/water partition coefficient (kow). The LogPow data were obtained from two different sources [35, 36]. The average LogPow of the most sensitive pesticides in SFC was 2.87. On the other hand, 3.27 was the average LogPow for those of LC. Considering a value of 3 as an average LogPow for those pesticides evaluated, the percentage of compounds below that value was 46% in LC and 44% in SFC. In the light of the results, logPow does not seem to be a parameter that influences sensitivity in one type of chromatography or another. Methomyl (0.09) and oxasulfuron (–0.81) were the pesticides with the lowest LogPow in LC and SFC, respectively. Bifenthrin was the pesticide with the highest LogPow in LC (6.60), while tau-fluvalinate was the pesticide with the highest LogPow in SFC (7.02).
Ion suppression
Matrix effects were evaluated through two different experiments. First, matrix effects were calculated by comparing the slope values of the calibration curves in solvent with those values of the matrix-matched calibration curves [37]. Then, the total ion chromatograms (TIC) of each matrix were compared with the chromatograms of the corresponding multiresidue method. The aim of these experiments was to determine whether matrix ion suppression was also influenced by the different elution profiles in each chromatography. Matrix effects between 0 and 20% were considered low or non-existent; however, modifications of the signal between 20 and 50% and >50% were considered as medium and strong matrix effects, respectively. As exposed in Table 1, there were no big differences between types of chromatography in tomato matrix, with the results being slightly better in LC. However, focusing on the leek matrix, there was a huge difference in the percentage of compounds with a low or non-existent matrix effect, 5% and 28% for LC and SFC, respectively. Furthermore, the strong matrix effects in LC were 22% higher in comparison with SFC. As can be observed in the Table 1, the results in onion matrix were very similar to those in leek. Regarding the orange matrix, this difference was even more significant: 7% of compounds in LC had matrix effects below 20% while 53% of the pesticides showed low matrix effects in SFC. Our previous studies denoted that SFC analysis provides in general terms a low matrix effect due to its nebulization and sampling efficiency in the ion source [22, 38]. These data confirm that SFC is effective in reducing matrix effects in matrices with a high number of interfering components like onion, leek, and orange. However, in matrices like tomato, there were no vast differences.
To evaluate the performance of both chromatographies against ion suppression, an injection of a blank extract of each matrix was carried out scanning the sample with the quadrupoles Q1 and Q3 and analyzing in the range of 100 to 1000 m/z (positive and negative polarity). The total ion chromatograms (TIC) were then overlapped with the chromatograms of the multiresidue methods. The analytes overlap** with the TIC of tomato, onion, and leek were challenging to evaluate. Based on the intensity of the TICs in matrix, the interactions between the analytes and the interferents of tomato, onion, and leek were similar in both methods. Tomato TIC intensity was low compared to that of the other two matrices, and no strong interference of the TIC with the compounds was observed. The opposite situation was found in leek and onion matrices, where a huge TIC intensity overlapped most of the pesticide’s signals in both techniques. However, the most interesting case involved the orange matrix, where the matrix effect differences were higher. In Figure 3, both TIC and multiresidue chromatograms have been equally scaled. At first sight, one may think that matrix effects should be higher in SFC due to the high TIC intensity. However, 75% of SFC compounds eluted before minute 5 (the last compound eluted at 7.622 min), and the intensity of the TIC in that region is approximately half compared to the compound elution zone in the LC chromatogram. This approach implies that the reduction of matrix effects in SFC is not only related to its sampling efficiency in the ESI source. The different elution mechanisms between technics play an important role in matrix effects as well. The data of identified compounds (see subsection “Ion source temperature evaluation”) and matrix effects in leek and onion showed better results in SFC compared to LC. Furthermore, in orange, this enhancement was even stronger. For example, azinphos-ethyl, diniconazole, fenpropathrin, and propiconazole were not identified at 350 °C in leek and orange matrices using LC. Regarding the SFC results for these pesticides, they were not identified in leek, but they were identified in orange matrix. All these compounds have retention times ranging between minutes 1 and 4, which means that they eluted with the lowest amount of coextracted matrix compounds in SFC. This evidence suggests that the good results of the orange matrix were thanks to a combination of the ionization efficiency and reduction of ion suppression due to the decrease of coelution with matrix components.
Conclusions
A comparison of two types of chromatography has been made using the same ESI-MS platform. Different ion source temperatures were studied: LC needed the highest temperature tested (350 °C) to achieve the highest sensitivity. However, in SFC, the sensitivity differences were much lower between the higher and the lower temperatures tested. In terms of sensitivity, SFC showed better results in complex matrices like leek, onion, and orange. Focusing on the compounds that presented higher sensitivity in all the studied matrices in each technic, it was observed that triazoles were the chemical group whose sensitivity was higher in SFC, followed by organophosphates. In LC, a majority correspond to organophosphates. The most sensitive pesticides in each technic were studied in terms of their octanol/water partition coefficient and the values were similar in both cases, concluding that this parameter is not decisive in the prediction of which chromatography is the best for each pesticide. Matrix effects were evaluated in terms of ion suppression. Higher ion suppression was observed in LC when complex matrices were analyzed. Furthermore, a test comparing the TIC of the matrices proved that the different elution mechanisms play an important factor in the reduction of ion suppression. Considering the results, SFC provides a better performance regarding LC when the coextracts from the matrix are very high. Furthermore, the SFC advantages are also of interest when low ion source temperatures are necessary for thermolabile compound analysis.
References
Stachniuk A, Fornal E Liquid chromatography-mass spectrometry in the analysis of pesticide residues in food. Food Anal Methods (2016).
Masiá A, Suarez-Varela MM, Llopis-Gonzalez A, Picó Y Determination of pesticides and veterinary drug residues in food by liquid chromatography-mass spectrometry: a review. Anal Chim Acta (2016).
McClain R. Milestones in supercritical fluid chromatography: Supercritical Fluid Chromatography; 2017.
Pilařová V, Plachká K, Khalikova MA, Svec F, Nováková L. Recent developments in supercritical fluid chromatography – mass spectrometry: is it a viable option for analysis of complex samples? Chem: TrAC - Trends Anal; 2019.
Grand-Guillaume Perrenoud A, Veuthey JL, Guillarme D. Comparison of ultra-high performance supercritical fluid chromatography and ultra-high performance liquid chromatography for the analysis of pharmaceutical compounds. J Chromatogr A. 2012. https://doi.org/10.1016/j.chroma.2012.10.005.
Saito M, Yamauchi Y, Kashiwazaki H, Sugawara M. New pressure regulating system for constant mass flow supercritical-fluid chromatography and physico-chemical analysis of mass-flow reduction in pressure programming by analogous circuit model. Chromatographia. 1988;25:801–5. https://doi.org/10.1007/BF02262088.
Guillarme D, Desfontaine V, Heinisch S, Veuthey JL What are the current solutions for interfacing supercritical fluid chromatography and mass spectrometry? J. Chromatogr. B Anal. Technol. Biomed. Life Sci. (2018).
West C, Melin J, Ansouri H, Mengue Metogo M. Unravelling the effects of mobile phase additives in supercritical fluid chromatography Part I: Polarity and acidity of the mobile phase. J Chromatogr A. doi. 2017. https://doi.org/10.1016/j.chroma.2017.02.066.
West C, Lemasson E. Unravelling the effects of mobile phase additives in supercritical fluid chromatography—part II: adsorption on the stationary phase. J Chromatogr A. 2019. https://doi.org/10.1016/j.chroma.2019.02.002.
West C, Lemasson E, Bertin S, Hennig P, Lesellier E. An improved classification of stationary phases for ultra-high performance supercritical fluid chromatography. J Chromatogr A. 2016. https://doi.org/10.1016/j.chroma.2016.02.052.
Giuffrida D, Zoccali M, Mondello L. Carotenoid and apocarotenoid analysis by SFE-SFC-QqQ/MS. Methods Mol Biol. 2020. https://doi.org/10.1007/978-1-4939-9952-1_16.
Wicker AP, Carlton DD, Tanaka K, Nishimura M, Chen V, Ogura T, et al. On-line supercritical fluid extraction—supercritical fluid chromatography-mass spectrometry of polycyclic aromatic hydrocarbons in soil. J Chromatogr B Anal Technol Biomed Life Sci. 2018. https://doi.org/10.1016/j.jchromb.2018.04.014.
Schulze S, Paschke H, Meier T, Muschket M, Reemtsma T, Berger U. A rapid method for quantification of persistent and mobile organic substances in water using supercritical fluid chromatography coupled to high-resolution mass spectrometry. Anal Bioanal Chem. 2020. https://doi.org/10.1007/s00216-020-02722-5.
Losacco GL, Ismail O, Pezzatti J, González-Ruiz V, Boccard J, Rudaz S, et al. Applicability of supercritical fluid chromatography–mass spectrometry to metabolomics II–Assessment of a comprehensive library of metabolites and evaluation of biological matrices. J Chromatogr A. 2020. https://doi.org/10.1016/j.chroma.2020.461021.
West C. Recent trends in chiral supercritical fluid chromatography. TrAC Trends Anal Chem. 2019;120:115648. https://doi.org/10.1016/j.trac.2019.115648.
Felletti S, Ismail OH, De Luca C, Costa V, Gasparrini F, Pasti L, et al. Recent achievements and future challenges in supercritical fluid chromatography for the enantioselective separation of chiral pharmaceuticals. Chromatographia. 2019;82:65–75. https://doi.org/10.1007/s10337-018-3606-1.
Speybrouck D, Lipka E. Preparative supercritical fluid chromatography: a powerful tool for chiral separations. J Chromatogr A. 2016;1467:33–55. https://doi.org/10.1016/j.chroma.2016.07.050.
Bernal J, Ares AM, Toribio L Chapter 17 - Applications to food analysis. In: Poole CFBT-SFC (ed). Elsevier (2017) 495–514.
Ishibashi M, Ando T, Sakai M, Matsubara A, Uchikata T, Fukusaki E, et al. High-throughput simultaneous analysis of pesticides by supercritical fluid chromatography/tandem mass spectrometry. J Chromatogr A. 2012;1266:143–8. https://doi.org/10.1016/j.chroma.2012.09.067.
Cutillas V, Galera MM, Rajski Ł, Fernández-Alba AR. Evaluation of supercritical fluid chromatography coupled to tandem mass spectrometry for pesticide residues in food. J Chromatogr A. 2018:1545. https://doi.org/10.1016/j.chroma.2018.02.048.
Ishibashi M, Izumi Y, Sakai M, Ando T, Fukusaki E, Bamba T. High-throughput simultaneous analysis of pesticides by supercritical fluid chromatography coupled with high-resolution mass spectrometry. J Agric Food Chem. 2015;63:4457–63. https://doi.org/10.1021/jf5056248.
Pérez-Mayán L, Cobo-Golpe M, Ramil M, Cela R, Rodríguez I. Evaluation of supercritical fluid chromatography accurate mass spectrometry for neonicotinoid compounds determination in wine samples. J Chromatogr A. 1620;2020:460963. https://doi.org/10.1016/j.chroma.2020.460963.
Tan Q, Fan J, Gao R, He R, Wang T, Zhang Y, et al. Stereoselective quantification of triticonazole in vegetables by supercritical fluid chromatography. Talanta. 2017;164:362–7. https://doi.org/10.1016/j.talanta.2016.08.077.
Cutillas V, García-Valverde M, Gómez-Ramos M, del M, Díaz-Galiano FJ, Ferrer C, et al. Supercritical fluid chromatography separation of chiral pesticides: unique capabilities to study cyhalothrin and metalaxyl as examples. J Chromatogr A. 1620;2020:461007. https://doi.org/10.1016/j.chroma.2020.461007.
Zhao L, **e J, Guo F, Liu K. Enantioseparation of napropamide by supercritical fluid chromatography: effects of the chromatographic conditions and separation mechanism. Chirality. 2018. https://doi.org/10.1002/chir.22836.
Anastassiades M, Lehotay SJ, Štajnbaher D, Schenck FJ. Fast and easy multiresidue method employing acetonitrile extraction/partitioning and “dispersive solid-phase extraction” for the determination of pesticide residues in produce. J AOAC Int. 2003;86:412–31. https://doi.org/10.1093/jaoac/86.2.412.
SANTE/12682/2019 Method validation and quality control procedures for pesticide residues analysis in food and feed: Document No. SANTE/12682/2019 (2020) (Implemented by 01.01.2020)
Page JS, Kelly RT, Tang K, Smith RD. Ionization and transmission efficiency in an electrospray ionization–mass spectrometry interface. J Am Soc Mass Spectrom. 2007;18:1582–90. https://doi.org/10.1016/j.jasms.2007.05.018.
Janusson E, Hesketh AV, Bamford KL, Hatlelid K, Higgins R, McIndoe JS. Spatial effects on electrospray ionization response. Int J Mass Spectrom. 2015. https://doi.org/10.1016/j.ijms.2015.07.016.
Kruve A. Influence of mobile phase, source parameters and source type on electrospray ionization efficiency in negative ion mode. J Mass Spectrom. 2016. https://doi.org/10.1002/jms.3790.
Cutillas V, Jesús F, Ferrer C, Fernández-Alba AR. Overcoming difficulties in the evaluation of captan and folpet residues by supercritical fluid chromatography coupled to mass spectrometry. Talanta. 2021:223. https://doi.org/10.1016/j.talanta.2020.121714.
Rahman MM, Hiraoka K, Chen LC. Realizing nano electrospray ionization using disposable pipette tips under super atmospheric pressure. Analyst. 2014. https://doi.org/10.1039/c3an01635h.
Kirby AE, Jebrail MJ, Yang H, Wheeler AR. Folded emitters for nanoelectrospray ionization mass spectrometry. Rapid Commun Mass Spectrom. 2010. https://doi.org/10.1002/rcm.4787.
Lesellier E. Retention mechanisms in super/subcritical fluid chromatography on packed columns. J. Chromatogr. A. 2009.
EU Reference Laboratories (EURLs) for Residues of Pesticides (2006) EURL DataPool website. https://www.eurl-pesticides-datapool.eu/. Accessed 22 Mar 2021
IUPAC. Pesticide Properties Database: Int. Union Pure Appl; 2007.
Del Mar G-RM, Rajski Ł, Lozano A, Fernández-Alba AR. The evaluation of matrix effects in pesticide multi-residue methods: via matrix fingerprinting using liquid chromatography electrospray high-resolution mass spectrometry. Anal Methods. 2016;8:4664–73. https://doi.org/10.1039/c6ay00436a.
Cutillas V, Murcia-Morales M, Gómez-Ramos MDM, Taha SM, Fernández-Alba AR. Supercritical fluid chromatography coupled to tandem mass spectrometry for the analysis of pesticide residues in dried spices. Benefits and drawbacks. Anal Chim Acta. 2019:1059. https://doi.org/10.1016/j.aca.2019.01.010.
Acknowledgements
The authors would also like to thank Shimadzu Corporation for providing the equipment and constant support of the work.
Funding
The authors received funding support provided by the European Commission, DG SANTE (Grant decision SI2.802063). The authors wish to thank too, the SpanishGovernment for the financial support given to the Project “EQC2019-006678-P”through FEDER funds.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
ESM 1
(DOCX 48 kb)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Cutillas, V., Ferrer, C. & R. Fernández-Alba, A. Liquid chromatography versus supercritical fluid chromatography coupled to mass spectrometry: a comparative study of performance for multiresidue analysis of pesticides. Anal Bioanal Chem 413, 5849–5857 (2021). https://doi.org/10.1007/s00216-021-03565-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-021-03565-4