Introduction

Atrial fibrillation (AF) was first recognized as a risk factor for systemic embolism, especially stroke, in patients with rheumatic valvular disease [1,2,3,4,5]. Decades later, the Framingham study demonstrated that patients with non-valvular AF (NVAF) also had a greater risk of stroke than those without AF [1, 2, 6,7,8]. Transesophageal echocardiography (TEE) studies have since revealed that more than 10% of patients with NVAF exhibit left atrial thrombus, more than 90% of cases of which are located in the left atrial appendage (LAA) [9]. Based on these findings, percutaneous LAA occlusion has been investigated and has proven to be safe and effective for stroke prevention in patients with NVAF [1, 2, 10,11,12,13,14,15,16].

To date, more than 10 commercially available LAA occlusion devices have been approved in North America, Europe, and Asia [10, 17]. The most widely used are the plug type, including WATCHMAN (Boston Scientific), and the pacifier type, including Amulet (Abbott Vascular), Ultraseal LAA Occluder (Cardia), and LAmbre (Lifetech) [4, 17]. Many of the occlusion devices mentioned above are designed for single-lobe LAA and only come in limited sizes; thus, they are limited in their use for occlusion of LAA with a complicated anatomy, especially LAA with multiple lobes [17]. In contrast, the LAmbre occluder, which is a pacifier-type device, consists of an inner umbrella and an outer cover. The LAmbre occluder comes in 11 regular sizes for single-lobe LAA and 6 special sizes with a small umbrella and a large cover for multi-lobe LAA [17]. LAA closure with the LAmbre device showed encouraging results in previous studies [15, 18,19,20]. Surprisingly, the small-umbrella LAmbre device was underused in those studies, and there are only a couple of case reports showing its effectiveness [21,22,23]. In this study, we aimed to evaluate the feasibility, safety, and short-term efficacy of the small-umbrella LAmbre device for occlusion of morphologically complicated LAA in patients with NVAF.

Methods

Study population

This was a retrospective multi-center study. Between September 2017 and May 2021, 129 patients at three hospitals in China underwent LAA occlusion using the LAmbre device, and the small-umbrella LAmbre device was used in 30 of these patients.

The primary enrollment criteria included: (1) NVAF; (2) age ≥ 18 years; and (3) CHA2DS2-VASc score ≥ 2 (≥ 3 for females). Patients were also required to meet at least one of the following three conditions: (1) not suitable for long-term anticoagulation (contraindications to anticoagulation or documented poor adherence to anticoagulation and refusal to undergo anticoagulation even after personal and detailed advice); (2) stroke events when taking adequate anticoagulation therapy with evidence of thrombosis originating from the LAA; (3) HAS-BLED score ≥ 3.

Morphologically complicated LAA was defined as one of the following: (1) multi-lobe LAA; (2) single-lobe LAA with a small tubular body and a large orifice (see more in the Discussion Section).

All patients underwent transthoracic echocardiography and TEE to exclude intracardiac thrombosis, including LAA thrombosis 48 h before LAA occlusion [10]. Further exclusion criteria included other comorbidities that required anticoagulation, other comorbidities that required cardiac surgery, acute myocardial infarction or unstable angina, symptomatic carotid disease, hemorrhagic disease, presence of a prosthetic valve, severe valvular disease, left ventricular ejection fraction < 30%, New York Heart Association functional class IV, ≥ moderate pericardial effusion, estimated survival < 1 year, recent stroke or transient ischemic attack within 30 days, pregnancy, and infective endocarditis.

This study was performed in accordance with the Declaration of Helsinki and was approved by Ethics Committee of Fuwai Hospital (Approval Number: 2021–1575) on 06 December 2021. Informed consent was obtained from all participants.

Device implantation

The LAmbre LAA occlusion system and the implantation procedure have been described in previous studies [15, 20]. Implantation was guided by angiography and TEE under general anesthesia or by intracardiac echocardiography (ICE) under local anesthesia. For the regular-sized LAmbre device, the diameter of the cover was 4–6 mm greater than the umbrella (Fig. 1A and C). By contrast, the diameter of the cover was 12–14 mm greater than the umbrella for the small-umbrella device (Fig. 1B and D). Size selection was based on the anatomy of the LAA (angiographic measurements), and a final decision was made after discussion between at least two experienced operators.

Fig. 1
figure 1

The LAmbre LAA occlusion device. The schematic diagrams (A and B) and real-life products (C and D) of the LAmbre occlusion device. The upper part is the inner umbrella anchored inside the LAA. The lower part is the outer cover placed on the LAA orifice. The umbrella and the cover are connected by a short, flexible central waist. A and C The regular-sized device with a cover diameter 4–6 mm greater than the umbrella. B and D The small-umbrella device with a small umbrella and a large cover (diameter difference of 12–14 mm). LAA: left atrial appendage

Major periprocedural complications included death, pericardial effusion requiring intervention, stroke, major bleeding, device dislocation requiring surgery, and access vessel complications requiring intervention.

Treatment after LAA occlusion

After the procedure, patients underwent anticoagulation within 3 months, dual antiplatelet therapy at 3–6 months, and long-term mono-antiplatelet therapy 6 months after the procedure. The course of anticoagulation was extended in patients with stroke events or thrombosis formation on the device.

Follow-up

Three months after the procedure, a repeat TEE was performed. At 6 months, 1 year, and every year thereafter, follow-up contrast cardiac computed tomography (CT) or TEE was performed in cases of suspected thrombosis.

The primary efficacy endpoint was successful device implantation and successful sealing of the LAA, as measured by TEE at 3 months after the procedure. Successful sealing was defined as no peri-device residual flow or < 3 mm of residual flow, as measured by TEE at 3 months after the procedure.

Major adverse events during follow-up included all-cause death, cardiovascular death, unexplained death, stroke, systemic embolism, device thrombosis, device dislocation, and serious pericardial effusion.

Statistical analysis

Continuous variables are expressed as mean ± standard deviation or median (interquartile range), as appropriate, and categorical parameters are expressed as ratio or percentage. Data analyses were performed using R version 4.0.2.

Results

Baseline characteristics of patients

The baseline clinical characteristics of all 30 patients are presented in Table 1. The mean age of patients was 66.8 ± 9.9 years, and 20 patients (66.7%) were male. Twenty-six patients had persistent AF (86.7%), and 4 patients had paroxysmal AF (13.3%). The most prominent comorbidity was hypertension (76.7%), and half of the patients had a history of transient ischemic attack or stroke. Heart failure (36.7%), diabetes mellitus (30%), and coronary artery disease (26.7%) were also common comorbid conditions. The mean CHA2DS2-VASc and HAS-BLED scores were 4 ± 1.6 and 3 ± 1.0, respectively. The mean left atrial diameter was 42 ± 6.7 mm, and the mean left ventricular ejection fraction was 58 ± 9.4%.

Table 1 Baseline characteristics of all patients (n = 30)

Procedure details

The small-umbrella LAmbre device (Fig. 1B and D) was successfully implanted in all 30 patients. A step-by-step explanation of the procedure is shown in Fig. 2. The procedural details are shown in Table 2. Two thirds of LAAs were cauliflower-type. Nearly three-quarters of patients (73.3%) had at least two lobes in the LAA. The inner umbrella of the device was anchored in the branch in 9 patients (30%), in the common trunks of branches in 13 patients, and in the main lobe in 8 patients with single-lobe LAA. The diameter of the LAA orifice was much greater than the LAA landing zone (27.4 ± 3.95 mm vs. 19.0 ± 4.39 mm, respectively) (Table 2). Accordingly, the mean diameters of the outer cover and inner umbrella of the occluder were 34.4 ± 2.75 mm and 22.0 ± 3.42 mm, respectively. Three of the 22 cases in which devices were anchored in the branch or common trunk are shown in Fig. 3. The distribution of different-sized occluders is shown in Fig. 4. The most frequently used sizes were 22/34 mm and 26/38 mm.

Fig. 2
figure 2

A step-by-step illustration of the implantation procedure. A LAA angiography. B Umbrella deployment. C Cover deployment. D Complete sealing of the LAA before release. E Tug test before final release. F Complete sealing of the LAA after final release of the device. G and H Intracardiac echocardiography showing LAA occlusion by the LAmbre device. I Intracardiac echocardiography showing no peri-device leak. LAA: left atrial appendage; X1: diameter of the landing zone; X2: length of the central flexible waist; X3: diameter of the LAA orifice

Table 2 Procedure details
Fig. 3
figure 3

The small-umbrella LAmbre device anchored in the branch for multi-lobe LAA. A–D Case 1 using a small-umbrella LAmbre device. E–H Case 2 using a small-umbrella LAmbre device. I–L Case 3 using a small-umbrella LAmbre device. A, E, I LAA angiography. B, F, J Complete sealing of the LAA by the device. C, G, K Intracardiac echocardiography showing LAA occlusion by the LAmbre device. D, H, L Intracardiac echocardiography showing no peri-device leak. LAA: left atrial appendage; X1: diameter of the landing zone; X2: length of the central flexible waist; X3: diameter of the LAA orifice

Fig. 4
figure 4

The size distribution of LAmbre devices

Implantation was successfully performed with the first selected device in 21 patients (70.0%), and satisfactory occlusion was achieved at the first attempt in 14 patients (46.7%) (Table 2). The procedure was guided by TEE in 18 patients (60%) and by ICE in 12 patients (40%). No peri-device residual flow was detected in 26 patients (86.7%), and mild (< 1 mm) residual flow was detected in 4 patients (13.3%) after final release of the occluder (Table 2). There were no serious periprocedural complications (Additional file 1: Table 1). Mild pericardial effusion (≤ 3 mm) occurred in 3 patients, which disappeared within 48 h after the procedure, and none of them needed further interventions.

Follow-up

At 3 months after the procedure, follow-up TEE was performed in 24 patients. No cases of peri-device residual flow were observed (Additional file 1: Table 2). Device thrombosis detected by TEE was reported in 1 patient at the 3-month follow-up. The thrombosis was on the surface of the device. The patient was treated with dabigatran after the procedure, and it was continued for another 3 months. At the 6-month follow-up by TEE, the thrombosis disappeared. Ischemic stroke occurred in 1 patient 1 month after the procedure. The functional deficit was mild with a National Institutes of Health Stroke Scale of 3. No thrombolysis was needed. This patient also had a prior history of stroke. No stroke events occurred during the following 1-year follow-up for this patient. At the 6-month and 1-year follow-up of all patients by cardiac CT, no peri-device residual flow or device-related thrombosis was reported. No death or hemorrhagic stroke events were reported during follow-up.

Within 3 months after the procedure, 4 patients were treated with warfarin, 5 with dabigatran, and 21 with rivaroxaban. Except for the 2 patients mentioned above, other 28 patients were treated with aspirin and clopidogrel from the third to sixth months after the procedure. All patients received long-term mono-antiplatelet therapy with aspirin 6 months after the procedure.

Discussion

To our knowledge, this study is the first to explore the feasibility, safety, and efficacy of the small-umbrella LAmbre device for occlusion of morphologically complicated LAA. The results demonstrate that LAA occlusion using this occluder was feasible with no major procedural complications or short-term adverse events in this study. The high success rates of implantation and occlusion during the procedure and at the short-term follow-up indicate that this occluder could be easily adaptable to various types of LAA, in particular anatomically complex LAA.

All patients in this study were strictly screened for LAA occlusion indications according to current guidelines and expert consensus [1, 2, 10]. All patients in this study were at a high risk of stroke, 15 (50%) of whom had a history of stroke (2 had previously experienced stroke more than twice), and 7 of whom had a history of confirmed or suspicious LAA thrombosis. The CHA2DS2-VASc score was ≥ 2 (≥ 3 for females) in all patients, ≥ 3 in 23 patients (76.7%), and ≥ 5 in 13 patients (43.3%). In summary, 20 patients underwent LAA occlusion for a high bleeding risk (HAS-BLED score of ≥ 3), and 2 of them also experienced stroke events when taking adequate anticoagulation therapy with evidence of thrombosis originating from the LAA. The other 10 patients had documented poor adherence to anticoagulation and refused anticoagulation, even after personal and detailed advice (5 of these patients had a history of stroke).

Unique features of the small-umbrella LAmbre device

Currently, most approved LAA occlusion devices are designed for single-lobe LAA or anatomically regular LAA. However, an autopsy study showed that LAAs have diverse morphologies, and up to 80% of LAAs might have two or more lobes [24]. The LAmbre LAA occlusion device comes in 17 sizes, which makes this device highly adaptable to different LAA morphologies, and they can be quite useful for occlusion of LAA with a complex anatomy. The small-umbrella LAmbre device was mainly designed for multi-lobe LAA. The ideal application was to implant the inner umbrella in one of the branches and place the outer cover on the orifice of the LAA (Fig. 5A). However, from our clinical experience, we believe that this small-umbrella LAmbre device could be used mainly in three conditions: (1) multiple-proximal-lobe LAA (Fig. 5A); (2) multiple-distal-lobe LAA with a large orifice (Fig. 5B); and (3) single-lobe LAA with a large orifice (Fig. 5C). In these conditions, the landing zone of the umbrella could be the branch (Fig. 5A), the common trunk (for multi-lobe LAA) (Fig. 5B), or the main lobe (for single-lobe LAA) (Fig. 5C). These LAAs share one common feature, which is the relatively large orifice compared with the lobe or the common trunk. In this study, the mean diameters of the LAA orifice and landing zone were 27.4 mm and 19.0 mm, respectively, with a ratio of 1.6, which was significantly greater than that in previous studies (23.6/22.7 mm in the study by Huang et al. [15], and 22.7/21.1 mm in the study by Park et al. [20]. Hence, in these patients, a cover that was relatively larger than the umbrella was needed. In this study, the difference in the diameter between the cover and the umbrella was at least 12 mm (Fig. 4). This special design is the key element for appropriate and safe anchoring of the small umbrella in the landing zone and thorough sealing of the large cover on the LAA orifice. Therefore, this study proved that the small-umbrella LAmbre device could be well adapted in morphologically complicated LAAs (Fig. 5).

Fig. 5
figure 5

A schematic diagram of the application of the small-umbrella LAmbre device in various types of LAA. A Multiple-proximal-lobe LAA, with the inner umbrella anchored in the branch. This is the most ideal condition for the small-umbrella LAmbre device. B Multiple-distal-lobe LAA with a large orifice and the inner umbrella anchored in the common trunk. C Single-lobe LAA with a large orifice and the inner umbrella anchored in the main lobe. LAA: left atrial appendage

Theoretically, with multiple-distal-lobe LAA with a large orifice (Fig. 5B) and single-lobe LAA with a large orifice (Fig. 5C), larger plug-type LAA occlusion devices could also be used, while traditional pacifier-type devices with regular sizes might not be suitable.

Feasibility, safety, and short-term efficacy of the small-umbrella LAmbre device

Implantation was successful in all 30 patients, supporting the feasibility of the small-umbrella LAmbre device. Due to the relatively complicated anatomy of the LAA, 9 patients (30%) required device size reselection, and 16 patients (53.3%) required device repositioning. However, only 3 patients demonstrated pericardial effusion after the procedure, with no other major or minor procedural complications. Hence, the small-umbrella occluder was safe compared with regular-sized devices [15].

During the procedure, only 4 patients had minimal peri-device residual flow (< 1 mm), indicating a 100% acute sealing rate. Of the 24 patients examined by TEE 3 months after the procedure, none had peri-device residual flow, suggesting a high short-term sealing rate. Therefore, the small-umbrella LAmbre device effectively sealed morphologically complicated LAA.

LAA occlusion in 12 patients (40%) was guided by ICE. In these patients, the procedure could be performed under local anesthesia. ICE could also be advanced into the left atrium to obtain a better view of the LAA and the occlusion device than TEE. This could shorten the procedure time, reduce anesthesia-related complications, and guarantee a better sealing effect.

Adverse events during follow-up

One patient experienced stroke at 1 month after the procedure, although contrast-enhanced cardiac CT and TEE suggested that the LAA was perfectly sealed. This patient might have had non-cardiac reasons for stroke. One patient demonstrated thrombosis formation on the device at the 3-month TEE follow-up, and adequate anticoagulation was prescribed for 3 months. The thrombosis was resolved at the 6-month follow-up TEE. No other adverse events were reported, including death, hemorrhagic stroke, systemic thromboembolism, device dislocation, or serious pericardial effusion. Overall, the adverse event rate was low during short-term follow-up. The absence of tamponade and device dislodgement is due to the unique design of LAmbre, in which the umbrella behaves like a balloon expandible stent, and once fully expanded, no further radial force is continuously affecting the LAA wall.

Limitations

This study has several limitations. First, it was a retrospective study, and no control groups, such as groups in which other devices were used for LAA occlusion, were included. Second, the sample size was relatively small, which might be due to the strict enrollment criteria for LAA occlusion and the relatively low rate of anatomically complicated LAA in patients with indications. However, this is the largest study reporting the use of the small-umbrella LAmbre device for LAA occlusion. Third, this study focused on evaluating the peri-procedural and short-term safety and efficacy outcomes after LAmbre device implantation. A long-term follow-up study is currently ongoing. The favorable outcomes in the short-term and the special design of this device suggest a promising outcome in the long-term. Fourth, implantation of the small-umbrella LAmbre device designed for morphologically complicated LAA might require greater operator experience.

Conclusions

Occlusion of morphologically complicated LAA using the small-umbrella LAmbre device was feasible, safe, and effective in patients with NVAF in this study. The small-umbrella occluder provides an alternative for patients who are not optimal candidates for regular-sized LAA occlusion devices.