Abstract
Objectives
The susceptibility of CT imaging to metallic objects gives rise to strong streak artefacts and skewed information about the attenuation medium around the metallic implants. This metal-induced artefact in CT images leads to inaccurate attenuation correction in PET/CT imaging. This study investigates the potential of deep learning–based metal artefact reduction (MAR) in quantitative PET/CT imaging.
Methods
Deep learning–based metal artefact reduction approaches were implemented in the image (DLI-MAR) and projection (DLP-MAR) domains. The proposed algorithms were quantitatively compared to the normalized MAR (NMAR) method using simulated and clinical studies. Eighty metal-free CT images were employed for simulation of metal artefact as well as training and evaluation of the aforementioned MAR approaches. Thirty 18F-FDG PET/CT images affected by the presence of metallic implants were retrospectively employed for clinical assessment of the MAR techniques.
Results
The evaluation of MAR techniques on the simulation dataset demonstrated the superior performance of the DLI-MAR approach (structural similarity (SSIM) = 0.95 ± 0.2 compared to 0.94 ± 0.2 and 0.93 ± 0.3 obtained using DLP-MAR and NMAR, respectively) in minimizing metal artefacts in CT images. The presence of metallic artefacts in CT images or PET attenuation correction maps led to quantitative bias, image artefacts and under- and overestimation of scatter correction of PET images. The DLI-MAR technique led to a quantitative PET bias of 1.3 ± 3% compared to 10.5 ± 6% without MAR and 3.2 ± 0.5% achieved by NMAR.
Conclusion
The DLI-MAR technique was able to reduce the adverse effects of metal artefacts on PET images through the generation of accurate attenuation maps from corrupted CT images.
Key Points
• The presence of metallic objects, such as dental implants, gives rise to severe photon starvation, beam hardening and scattering, thus leading to adverse artefacts in reconstructed CT images.
• The aim of this work is to develop and evaluate a deep learning–based MAR to improve CT-based attenuation and scatter correction in PET/CT imaging.
• Deep learning–based MAR in the image (DLI-MAR) domain outperformed its counterpart implemented in the projection (DLP-MAR) domain. The DLI-MAR approach minimized the adverse impact of metal artefacts on whole-body PET images through generating accurate attenuation maps from corrupted CT images.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
A large number of patients referred for whole-body PET/CT examinations present with metallic objects, such as coiling, dental implants/filling and hip/shoulder prostheses. These highly attenuating metallic objects give rise to severe photon starvation, beam hardening and scattering, thus leading to adverse artefacts, including streak, star-shape and voids in the reconstructed CT images. The presence of metallic objects in CT imaging commonly leads to severe deterioration of image quality by causing strong bright and dark streak artefacts, not only in the site of the metallic object but also in the surrounding regions. Metal-induced artefacts could skew the CT signal (depending on the size and material of the metallic object) or generate artificial signals, which might adversely impact the interpretation of CT images and hence potentially clinical diagnosis. In addition to the adverse impact of metal-induced artefacts on the visual assessment of CT images, the quantitative accuracy of the CT signal in the vicinity of metallic objects could be affected [1]. The quantitative analysis of CT images in the vicinity of the metallic object revealed large under/overestimation of tissue densities due to the distortion of the CT signal [2].
In addition to diagnostic imaging applications with contrast enhancement of CT in radiology, this imaging modality is also employed in quantitative positron emission tomography (PET) imaging in the framework of CT-based attenuation correction (AC) of the PET data [3]. Any issues or image artefacts affecting CT images would be reflected in the attenuation-corrected PET images, potentially leading to misdiagnosis and/or quantitative errors [4, 5]. Moreover, CT images commonly serve as standard of reference for the evaluation of MRI-guided attenuation correction [6,7,8] and MRI-only treatment planning in radiation oncology [9].
The presence of metallic objects in PET imaging itself does not cause any signal loss/distortion; however, the incorrect CT values (due to metal artefacts) could lead to inaccurate PET attenuation and scatter corrections. Noticeable over/underestimation of the activity concentration would be observed in the vicinity of metallic objects. Moreover, strong streak artefacts in CT images may distort genuine PET signal and may also cause factitious signals in the PET image (e.g. false abnormalities). When metallic objects in CT-based PET attenuation map cause inaccurate scatter correction, its impact will not be confounded to the site of the metal artefact and would affect the entire PET volume [4]. Metal artefacts are easily detectable through visual inspection of CT images; however, radiotracer uptake bias in PET images affected by incorrect AC maps owing to metal artefacts is not easily recognized without reviewing the attenuation map. Hence, metal artefact correction in PET imaging is as important as in CT imaging.
To address this issue, a number of metal artefact reduction (MAR) approaches have been proposed in the literature [3, 10]. However, none of them was widely adopted as standard of reference, and, as such, MAR still remains one of the major challenges in CT and PET/CT imaging [11].
A state-of-the-art MAR approach based on the prior image is the normalized MAR (NMAR), wherein thresholding-based image/tissue classification is applied on the original/affected or linear interpolation (LI)–corrected image to generate a prior image [12]. This approach tends to extract as much as possible information from unaffected regions in the original or partly corrected images in order to predict/correct/estimate the missing or affected regions in the original image. Iterative image reconstruction approaches tend to suppress metal artefacts through reconstruction of CT images from the unaffected or preliminary corrected projections/bins [13,14,15,16]. Despite these efforts, it is challenging to achieve acceptable/satisfactory outcomes for all metal artefacts using a single MAR framework owing to the high variety of sizes, materials, locations and background structures of metal implants. In this light, several attempts have been made to combine two or three MAR methods to devise a hybrid approach [17]. Hybrid approaches have shown superior performance compared to each of the MAR techniques alone [3]. These methods tend to provide a more realistic/accurate modelling of metal artefacts by taking beam hardening and Poisson noise into account [13] or employing a preprocessing step (for instance using other MAR techniques), which results in a case-specific prior knowledge for the problem at hand [17]. Hence, these methods combine the prior knowledge and/or conventional MAR technique with an iterative image reconstruction algorithm fine-tuned for MAR.
Deep learning–based algorithms emerged as promising approaches for solving a variety of image analysis and pattern recognition problems and have been successfully implemented in the context of metal artefact reduction in CT imaging [18,19,20,21,22,23]. Deep learning–based MAR techniques could be categorized as the 4th type of MAR approaches, wherein the correction for metal-induced beam hardening is applied in either the projection [22] or image domain [19]. MAR methods based on beam hardening correction, such as NMAR, have limited capability in the presence of high attenuating metal implants. Deep learning–based MAR approaches have demonstrated the ability to refine the performance of the NMAR algorithm, either in the projection or image domain, to eliminate the residual errors or new artefacts introduced by NMAR [24].
The aim of this work is to develop and evaluate a deep learning–based MAR to improve CT-based attenuation and scatter correction in PET/CT imaging. To this end, CT images affected by metallic artefacts were processed by the deep learning–based MAR to assess the impact on the resulting PET images. The proposed approaches were also compared to the NMAR, serving as baseline technique for comparison.
Materials and methods
Data acquisition
Two clinical databases were utilized in this work, including a simulation dataset used for training and assessment of the deep learning–based MAR approaches and whole-body 18F-FDG PET/CT images used for clinical evaluation. The simulation dataset consisted of 80 metal-free whole-body CT images acquired on a Biograph mCT PET/CT scanner (Siemens Healthcare) using the following parameters: effective tube current = 100 mAs, tube voltage = 110–120 kVp, slice thickness = 3 mm, automated tube voltage selection, automated tube current modulation, pitch factor of 1, mean CTDIvol = 5.8 mGy and DLP of 65 mGy·cm (average BMI = 29.5). Eighty patients selected from whole-body CT scans with no metal artefact (metal artefact–free) were included in the simulation dataset for the training of the network.
The PET/CT dataset used for clinical evaluation consists of 30 patients (mean age ± SD = 60 ± 8 years and mean weight ± SD = 70 ± 8 kg) presenting with substantial dental fillings (24), hip (3) and shoulder joint prostheses (3). For clinical evaluation, 30 patients presenting with substantial dental fillings, hip and shoulder joint prostheses were selected from whole-body PET/CT scans. Whole-body PET/CT scans were performed on the same Biograph mCT scanner (using the same CT acquisition protocol) about 1 h post-injection of 250±45 MBq of 18F-FDG for an average acquisition time of 25±5 min. The study protocol was approved by Geneva Cantonal Research ethics committee.
Metal artefact simulation
The training of the deep learning–based MAR approaches involved the use of 80 metal-free CT images where metals were artificially inserted to create a metal artefact–affected dataset with known ground truth. These data contained metal-free, metal-inserted (ground truth images) and metal-affected CT images. Metal-free images are the original CT images whereas the metal-inserted images were generated by inserting metallic implants in the metal-free images. These metallic implants included dental fillings, spine fixation screws and hip and shoulder joint prostheses with linear attenuation coefficients of iron, copper, gold and titanium. The algorithm proposed in [19] was employed to simulate metallic artefacts in CT images (metal-affected images), wherein the metal-inserted images were segmented into soft tissue, bone and metal using a weighted thresholding-based approach [25] followed by assignment of the corresponding attenuation coefficients. Subsequently, the metal-affected CT images were reconstructed through a combination of forward projection and filtered backprojection approach. The adopted metal artefact simulation approach is briefly described below [19]. Further details about the simulation process are provided in Supplemental materials.
The simulation dataset consisted of 80 clinical CT images (45 men and 35 women, age 51 ± 12 years) with no metal implants. On this dataset, 8 different metal shapes were simulated using 4 different materials (namely iron, gold, titanium and copper) in the different regions of the body. For each case (shape and material), 25 metal artefact simulations were conducted resulting in 25 (realizations) × 8 (shapes) × 4 (materials) metal artefact–affected images/samples. Given the 800 simulated cases, 660 were employed for the training and 140 for the evaluation of the different methods, equally sampled from the different metal shapes, materials and anatomical regions. Mass attenuation coefficients of the metal inserts, bone and soft tissue were obtained from the XCOM photon cross section library [26]. The metal-free and metal-inserted images underwent a polychromatic Radon transform (following the algorithm described in [19]) and were reconstructed using a filtered backprojection (FBP) algorithm to generate standard of reference images (for evaluation of MAR techniques) and the metal-affected images, respectively. An X-ray source with 120 kVp tube potential and equi-angular fan-beam geometry was simulated. One thousand angular projections over 920 detector bins were calculated for each CT slice.
Metal artefact reduction strategies
Normalized metal artefact correction
The normalized MAR (NMAR) approach relies on prior knowledge obtained from a threshold-based image classification of the metal-affected or LI corrected image. In the first step, the metal implant is segmented from the input image followed by forward projection and the identification of the metal trace in the projection space. Prior knowledge is used to normalize the metal-affected data in the projection space before applying interpolation to the trace of the metal implant [12].
The NMAR approach was applied on both real (clinical) and simulated metal-artefacted images, providing a baseline for evaluation of the deep learning–based MAR approaches. Moreover, the outcome of the NMAR approach was employed as prior knowledge or auxiliary image to be fed into the deep learning model.
Deep learning–based MAR in the image domain
Deep learning–based MAR can be implemented in either the projection or image domains (or any combination of them). MAR in the image domain is considered as an end-to-end image transformation or regression from metal-affected to artefact-corrected (or artefact-free) CT images. To implement MAR in the image domain, three different strategies were explored: (i) the input image to the deep learning model is restricted to the metal-affected image to estimate the artefact-corrected CT image, (ii) the input image is the metal artefact–corrected CT image using the NMAR technique (CT-NMAR), (iii) the input to the deep learning model consist of both metal-affected and CT-NMAR images fed in two different channels (Fig. 1a). Visual inspection and preliminary quantitative analysis revealed the superior performance of the third scenario. Hence, in the rest of this work, we focus only on this approach.
Two-hundred metal-free, simulated metal-inserted and metal-affected volumetric CT images were employed to train the deep learning model in the image domain. To this end, a dilated convolutional neural network with residual connections (HighResNet) [27] was adopted and configured in the open-source NiftyNet deep learning platform [28]. Overall, 10,000 2D simulated metal-affected and metal-inserted CT images were used for the training whereas 2000 2D CT images were used for evaluation of the model. The model consisted of two entry channels of 400 × 400 voxels taking metal-affected and CT-NMAR images as input and one output channel of 400 × 400 voxels. The HighResNet network is composed of 20 residual convolutional layers regulated by 3 × 3 × 3 kernels dilated by factors two and four at the different layers (Fig. 1a). To train the model, an L2 loss function was adopted using the Adam optimization and learning rate from 0.06 to 0.01 following the procedure recommended in [29].
Deep learning–based MAR in the projection domain
To implement the deep learning–based MAR in the projection domain, the same three strategies explored in the image domain were adopted, except that image-to-image regression was carried on 2D CT projection data after applying the Radon transform. Contrary to the image domain, none of the three strategies exhibited superior performance. However, the first strategy, wherein the input to the deep learning model is the metal-affected projections to generate the artefact-free projections (Fig. 2b), was selected owing to its higher robustness and computational efficiency. The same deep learning model and training procedure described in the previous section were used to implement MAR in the projection domain. The resulting artefact-free projections were reconstructed using the FBP algorithm with 1000 angular projections over 920 detector bins.
Representative examples of metal artefacts simulation and correction showing a CT images after inserting metallic implants (standard of reference). b Metal-affected CT images. CT images after applying MAR using (c) NMAR, (d) deep learning algorithm operating in the projection domain (DLP-MAR) and (e) deep learning algorithm operating in the image domain (DLI-MAR)
MAR in PET/CT imaging
To evaluate the impact of MAR in whole-body PET/CT imaging, a clinical database consisting of 30 patients presenting with severe metal artefacts in CT images was collected. The metal implants are dental fillings (24), hips (3) and shoulder joint prostheses (3). The raw PET data were corrected for attenuation using the original CT images (before MAR), CT images corrected by NMAR algorithm and CT images corrected by the deep learning approaches. The quality and quantitative analysis of PET images corrected for attenuation using CT images before and after MAR was assessed to put into perspective the potential impact of MAR in PET imaging.
The PET raw data were reconstructed using 3D ordinary Poisson-ordered subset expectation maximization (OP-OSEM) algorithm with 2 iterations and 21 subsets in 400 × 400 matrix with a voxel size of 2 × 2 × 3 mm. CT-based scatter correction, time-of-flight (TOF) information and point spread function (PSF) modelling were considered during PET image reconstruction using the offline e7 tool (Siemens Healthcare).
Evaluation strategy
Evaluation of MAR techniques was carried out on CT and PET images corrected for attenuation using CT images before and after applying MAR. For quantitative evaluation of CT images, peak signal-to-noise ratio (PSNR), root mean square error (RMSE) and structural similarity (SSIM) index were calculated before and after MAR using Eqs. (1)–(3), respectively.
In Eq. (1), Pk indicates the maximum intensity of either CTref or CTMAR whereas MSE denotes the mean squared error between reference CT images (metal-inserted) and CT images after MAR. In Eq. (3), Averef and AveMAR are the mean values of CTref and CTMAR images, respectively. δref and δMAR stand for the variances and δref,MAR denotes the covariance of CTref and CTMAR images. The constant parameters (k1 = 0.01 and k2 = 0.02) were sought to avoid division by very small values. The metal-inserted CT images with no artefacts (generated by the simulation) were considered as reference.
Thirty clinical whole-body PET/CT images with noticeable metal artefacts were examined before and after applying MAR to CT images. The evaluation focused on artefacts induced on PET images by metal artefact present in the attenuation maps as well as PET quantification bias in the vicinity of the metallic implants before and after MAR. Moreover, the impact of MAR on scatter correction and consequently on PET quantification was assessed using whole-body 18F-FDG PET/CT studies. To this end, standardized uptake values (SUVs) were calculated within volumes of interest (VOIs) drawn on the affected regions before and after applying MAR using the different techniques. The VOIs were defined in such a way to cover a sufficiently large volume encompassing the entire metallic object, the surrounding affected/disturbed tissues and the residues of the metal artefacts. Larger VOIs were drawn for hip and shoulder joint prostheses with a maximum size of 950 cm3 whereas smaller VOIs were drawn for dental fillings with a minimum size of 55 cm3.
The significance of the differences between the different MAR techniques (quantitative metrics) was assessed using the paired t test method. P values smaller than 0.05 were regarded to reflect statistical significance.
Results
In the first part of the “Results” section, we report the evaluation of MAR techniques solely on CT images using the simulated metal artefacts dataset. Subsequently, the impact of metal artefacts (in CT images) and the different MAR techniques are assessed on the corresponding attenuation-corrected PET images.
Figure 2 (complete set of images is presented in Supplemental Figure 1) illustrates four examples of metal artefacts and the outcome of different MAR techniques from the simulation dataset. The metal-inserted images (artefact-free CT images inserted with metal implants) are regarded as standard of reference. Deep learning–based MAR in the image (DLI-MAR) and projection (DLP-MAR) domains were assessed along with NMAR techniques, which provides a yardstick for comparison. Visual inspection revealed the superior performance DLI-MAR compared to NMAR and DLP-MAR techniques.
Table 1 summarizes the performance metrics of the different MAR techniques on the simulation dataset. In agreement with visual inspection (Fig. 2), quantitative evaluation of the MAR techniques demonstrated the superior performance of DLI-MAR techniques. The p values in Table 1 were calculated between the results obtained from DLI-MAR and DLP-MAR (reported in the same table) techniques to examine the statistical significance of the differences between these two techniques.
The validation dataset consisted of 140 samples of simulated metal artefacts generated from the different metal shapes, materials and anatomical regions. A separate analysis of the results associated with a specific metal shape, material or anatomical region revealed no significant differences and variations among them. Hence, we concluded that the performance of the evaluated techniques is not correlated/linked to any specific anatomical region, metal shape or material. The average values over the entire validation cases are reported for each technique in Table 1. A similar observation was made in the clinical study wherein the different techniques performed similarly in the different regions of the body.
Although the evaluation of MAR techniques on CT images demonstrated the superior performance of DLI-MAR over DLP-MAR, there was no proof of statistically significant differences when the resulting CT images were used for PET attenuation correction. Therefore, only the results obtained from NMAR and DLI-MAR methods were presented when assessing the impact of MAR techniques on PET images.
Figures 3 and 4 present metal artefacts in CT images due to shoulder joint and hip prostheses, which gave rise to image artefacts in the corresponding PET images. The image artefacts are clearly visible around the site of metallic implants. MAR techniques, particularly DLI-MAR, eliminated noticeably the streak and dark bands from CT images, consequently reducing the artefacts around the site of the implants. The horizontal profiles drawn on PET images illustrate clearly the degree of PET signal alteration before MAR and recovery by the MAR techniques.
Representative example of image artefacts apparent on PET images from shoulder joint prosthesis in CT images. Original CT image before correction (a) and after MAR using NMAR (b) and DLI-MAR (c) methods. The corresponding attenuation-corrected PET images are displayed in d, e and f, respectively. Horizontal profiles drawn on the shoulder area of PET images show the impact of metal artefacts and their reduction on the resulting PET signals
Representative example of image artefacts apparent in the PET image from the hip prosthesis in CT images. Original CT image before correction (a) and after MAR using NMAR (b) and DLI-MAR (c) methods. The corresponding attenuation-corrected PET images are displayed in d, e and f, respectively. Horizontal profiles drawn on the pelvis area of PET images show the impact of metal artefacts on PET signals
Figures 5 and 6 depict two examples of impaired scatter estimation during PET image reconstruction owing to the presence of metallic artefacts in CT images. In Fig. 5, metal artefacts led to underestimation of the head contour due to the weak signals at the boundary of the head, which resulted in overestimation of the scatter fraction when using model-based scatter correction [30]. The scatter fraction for this bed position for PET-CTAC was 58% while scatter fractions of 33% and 24% were observed when using PET-NMAR and PET-DLI-MAR, respectively. Horizontal profiles drawn on PET images show the differences between PET images before and after applying the different MAR techniques.
Example of scatter correction overestimation due to metallic artefacts. Original CT image before correction (a) and after MAR using NMAR (b) and DLI-MAR (c) methods. The corresponding attenuation-corrected PET images are displayed in d, e and f, respectively. Horizontal profiles drawn on PET images compare the PET signals before and after application of MAR methods
Overestimation of tracer uptake (SUV) in a malignant lesion due to metallic artefacts. Original CT image before correction (a) and after MAR using NMAR (b) and DLI-MAR (c) methods. The corresponding attenuation-corrected PET images are displayed in d, e and f, respectively. Horizontal profiles drawn on the lesion on PET images compare the PET signals before and after application of MAR methods
Supplemental Figure 2 depicts an example wherein metal artefacts in CT images led to significant underestimation of scatter fraction during PET/CT image reconstruction. The scatter fraction in PET-CTAC was 11%, while scatter fractions of 19% and 25% were observed in PET-NMAR and PET-DLI-MAR images, respectively.
Figure 6 depicts an example wherein the lesion near the site of metal artefact was significantly overestimated in the PET-CTAC image. Application of MAR techniques on the CT image or attenuation map significantly reduced the seemingly artificial high SUV of the lesion.
Table 2 summarizes the quantitative impact of MAR techniques on PET images. Thirty VOIs were drawn at the site of metal artefacts on both clinical and simulation datasets. For the simulation dataset, the mean SUV bias (percentage of SUV under-/overestimation after MAR averaged across all patients) was measured within the VOIs considering the metal-inserted images (CT images with metal implant without artefacts) as standard of reference attenuation maps. For the clinical dataset, the bias (or change in tracer uptake) was calculated against the PET-CTAC before applying the MAR techniques owing to the lack of ground truth attenuation map. Overall, the simulation studies demonstrated that metallic artefacts in CT images led to significant overestimation of the SUV in the proximity of the metallic implant, while MAR techniques, in particular DLI-MAR, were able to reduce the adverse impact of the metallic artefacts on PET images. The large negative bias observed in the clinical studies confirmed the observations in the simulation dataset.
Discussion
Different strategies for deep learning–based metal artefact reduction (MAR) of CT images were explored either in the image or projection domains [ 18F-Fluorodeoxyglucose Convolutional neural networks Computed tomography Deep learning–based MAR in the image space Deep learning–based MAR in the projection space Linear Interpolation Metal artefact correction Normalized metal artefact correction Positron emission tomography Positron emission tomography/computed tomography Standardized uptake value Time of flight Volume of interest Whole body Feldhaus F, Böning G, Jonczyk M et al (2019) Metallic dental artifact reduction in computed tomography (Smart MAR): improvement of image quality and diagnostic confidence in patients with suspected head and neck pathology and oral implants. Eur J Radiol 118:153–160 Mehranian A, Ay MR, Rahmim A, Zaidi H (2013) X-ray CT metal artifact reduction using wavelet domain L0 sparse regularization. IEEE Trans Med Imaging 32:1707–1722 Abdoli M, Dierckx RAJO, Zaidi H (2012) Metal artifact reduction strategies for improved attenuation correction in hybrid PET/CT imaging. Med Phys 39:3343–3360 Lemmens C, Montandon M-L, Nuyts J, Ratib O, Dupont P, Zaidi H (2008) Impact of metal artefacts due to EEG electrodes in brain PET/CT imaging. Phys Med Biol 53:4417–4429 Abdoli M, de Jong JR, Pruim J, Dierckx RA, Zaidi H (2011) Reduction of artefacts caused by hip implants in CT-based attenuation-corrected PET images using 2-D interpolation of a virtual sinogram on an irregular grid. Eur J Nucl Med Mol Imaging 38:2257–2268 Arabi H, Zeng G, Zheng G, Zaidi H (2019) Novel adversarial semantic structure deep learning for MRI-guided attenuation correction in brain PET/MRI. Eur J Nucl Med Mol Imaging 46:2746–2759 Arabi H, Zaidi H (2017) Comparison of atlas-based techniques for whole-body bone segmentation. Med Image Anal 36:98–112 Arabi H, Zaidi H (2020) Truncation compensation and metallic dental implant artefact reduction in PET/MRI attenuation correction using deep learning-based object completion. Phys Med Biol 65:195002 Arabi H, Dowling JA, Burgos N et al (2018) Comparative study of algorithms for synthetic CT generation from MRI: consequences for MRI-guided radiation planning in the pelvic region. Med Phys 45:5218–5233 Abdoli M, Mehranian A, Ailianou A, Becker M, Zaidi H (2016) Assessment of metal artifact reduction methods in pelvic CT. Med Phys 43:1588–1597 Gjesteby L, Man BD, ** Y et al (2016) Metal artifact reduction in CT: where are we after four decades? IEEE Access 4:5826–5849 Meyer E, Raupach R, Lell M, Schmidt B, Kachelriess M (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37:5482–5493 Zhang X, Wang J, **ng L (2011) Metal artifact reduction in x-ray computed tomography (CT) by constrained optimization. Med Phys 38:701–711 De Man B, Nuyts J, Dupont P, Marchal G, Suetens P (2000) Reduction of metal streak artifacts in X-ray computed tomography using a transmission maximum a posteriori algorithm. IEEE Trans Nucl Sci 47:977–981 Lemmens C, Faul D, Nuyts J (2009) Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion. IEEE Trans Med Imaging 28:250–260 Laukamp KR, Zopfs D, Lennartz S et al (2019) Metal artifacts in patients with large dental implants and bridges: combination of metal artifact reduction algorithms and virtual monoenergetic images provides an approach to handle even strongest artifacts. Eur Radiol 29:4228–4238 Zhang Y, Yan H, Jia X, Yang J, Jiang SB, Mou X (2013) A hybrid metal artifact reduction algorithm for x-ray CT. Med Phys 40:041910 Liang K, Zhang L, Yang H, Yang Y, Chen Z, **ng Y (2019) Metal artifact reduction for practical dental computed tomography by improving interpolation-based reconstruction with deep learning. Med Phys 46:e823–e834 Zhang Y, Yu H (2018) Convolutional neural network based metal artifact reduction in x-ray computed tomography. IEEE Trans Med Imaging 37:1370–1381 Lossau Née Elss T, Nickisch H, Wissel T, Morlock M, Grass M (2020) Learning metal artifact reduction in cardiac CT images with moving pacemakers. Med Image Anal 61:101655 Wang J, Noble JH, Dawant BM (2019) Metal artifact reduction for the segmentation of the intra cochlear anatomy in CT images of the ear with 3D-conditional GANs. Med Image Anal 58:101553 Park HS, Lee SM, Kim HP, Seo JK, Chung YE (2018) CT sinogram-consistency learning for metal-induced beam hardening correction. Med Phys 45:5376–5384 Arabi H, Zaidi H (2020) Applications of artificial intelligence and deep learning in molecular imaging and radiotherapy. Eur J Hybrid Imaging 4:1–23 Gjesteby L, Yang Q, ** Y et al (2017) Deep learning methods for CT image-domain metal artifact reduction. Proc. SPIE, Developments in X-Ray Tomography XI, Vol. 10391, p. 103910W Weeks KJ, Montana GS (1997) Three-dimensional applicator system for carcinoma of the uterine cervix. Int J Radiat Oncol Biol Phys 37:455–463 Berger M (2010) XCOM: photon cross sections database. http://www.nist.gov/pml/data/xcom/index.cfm. Accessed May 2020 Li W, Wang G, Fidon L, Ourselin S, Cardoso MJ, Vercauteren T (2017) On the compactness, efficiency, and representation of 3D convolutional networks: brain parcellation as a pretext task. International conference on information processing in medical imaging. Springer, pp 348-360 Gibson E, Li W, Sudre C et al (2018) NiftyNet: a deep-learning platform for medical imaging. Comput Methods Programs Biomed 158:113–122 Smith LN (2018) A disciplined approach to neural network hyper-parameters: part 1--learning rate, batch size, momentum, and weight decay. ar**v preprint ar**v:180309820 Watson CC (2000) New, faster, image-based scatter correction for 3D PET. IEEE Trans Nucl Sci 47:1587–1594 Mehranian A, Zaidi H (2015) Impact of time-of-flight PET on quantification errors in MR imaging-based attenuation correction. J Nucl Med 56:635–641 Arabi H, Rager O, Alem A, Varoquaux A, Becker M, Zaidi H (2015) Clinical assessment of MR-guided 3-class and 4-class attenuation correction in PET/MR. Mol Imaging Biol 17:264–276 Arabi H, Zaidi H (2016) Magnetic resonance imaging-guided attenuation correction in whole-body PET/MRI using a sorted atlas approach. Med Image Anal 31:1–15 Hegazy MAA, Cho MH, Cho MH, Lee SY (2019) U-net based metal segmentation on projection domain for metal artifact reduction in dental CT. Biomed Eng Lett 9:375–385 Martin O, Boos J, Aissa J et al (2020) Impact of different iterative metal artifact reduction (iMAR) algorithms on PET/CT attenuation correction after port implementation. Eur J Radiol 129:109065 Martin O, Aissa J, Boos J et al (2020) Impact of different metal artifact reduction techniques on attenuation correction in 18F-FDG PET/CT examinations. Br J Radiol 93:20190069 Zaidi H, Koral KF (2004) Scatter modelling and compensation in emission tomography. Eur J Nucl Med Mol Imaging 31:761–782 Arabi H, Zaidi H (2020) Deep learning-guided estimation of attenuation correction factors from time-of-flight PET emission data. Med Image Anal 64:101718 Burgos N, Thielemans K, Cardoso MJ et al (2014) Effect of scatter correction when comparing attenuation maps: application to brain PET/MR2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), pp 1-5 van der Vos CS, Arens AIJ, Hamill JJ et al (2017) Metal artifact reduction of CT scans to improve PET/CT. J Nucl Med 58:1867–1872 Open Access funding provided by Université de Genève. This work was supported by the Swiss National Science Foundation under grant SNFN 320030_176052 and the Private Foundation of Geneva University Hospitals under Grant RC-06-01. The scientific guarantor of this publication is Habib Zaidi. The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article. No complex statistical methods were necessary for this paper. Written informed consent was waived by the Institutional Review Board. Institutional Review Board approval was obtained. • Retrospective • Experimental • Multicenter study Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. (DOCX 4209 kb) Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. Arabi, H., Zaidi, H. Deep learning–based metal artefact reduction in PET/CT imaging.
Eur Radiol 31, 6384–6396 (2021). https://doi.org/10.1007/s00330-021-07709-z Received: Revised: Accepted: Published: Issue Date: DOI: https://doi.org/10.1007/s00330-021-07709-zAbbreviations
References
Funding
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
Conflict of interest
Statistics and biometry
Informed consent
Ethical approval
Methodology
Additional information
Publisher’s note
Supplementary Information
ESM 1
Rights and permissions
About this article
Cite this article
Keywords