Rapid assessment of left ventricular systolic function in a pacemaker clinic using a hand-carried ultrasound device

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Objective

Pacemaker recipients with left ventricular (LV) dysfunction are potential candidates for upgrades to implantable defibrillators or cardiac resynchronization devices. This study sought to determine if a hand-carried ultrasound (HCU) device could be used for rapid, inexpensive identification of LV dysfunction in a busy pacemaker clinic.

Materials and methods

Eighty patients undergoing routine pacemaker check were enrolled. Patients underwent HCU imaging in the sitting position during device interrogation, by an internist who had 20 h of didactic training and 20 practice examinations. LV dysfunction was defined as ejection fraction (EF) <40%. Patients also underwent echocardiography limited to EF assessment by a sonographer using a full-feature platform.

Results

The mean age was 75 ± 13 years; 49% were female. Coronary artery disease was present in 29%; 82% were NYHA class I or II. At the time of HCU imaging, 48% of patients were receiving RV pacing. HCU images were interpretable in 91% (73/80) and required 3.7 ± 0.9 min to complete. Based on the full-feature echo, LV dysfunction prevalence was 17/80 (21%); 25% of these patients were NYHA class I. The sensitivity of the HCU exam was 75%, specificity was 91%, negative predictive value was 93%, positive predictive value was 71%, and accuracy was 88%.

Conclusions

HCU screening in a pacemaker clinic by a non-cardiologist can rapidly and accurately identify pacemaker recipients with at least moderate LV dysfunction who might be candidates for device upgrades. Ventricular dyssynchrony associated with RV pacing does not limit HCU identification of LV dysfunction.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
EUR 32.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or Ebook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price includes VAT (Germany)

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R. et al., for the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators (2005). Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. New England Journal of Medicine, 352, 225–237.

    Article  PubMed  CAS  Google Scholar 

  2. Moss, A., Zareba, W., & Hall, W. et al. (2002). Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. For the Multicenter Automatic Defibrillator Implantation Trial II Investigators. New England Journal of Medicine, 346, 877–883.

    Article  PubMed  Google Scholar 

  3. Desai, A. S., Fang, J. C., Maisel, W. H., & Baughman, K. L. (2004). Implantable defibrillators for the prevention of mortality in patients with nonischemic cardiomyopathy: A meta-analysis of randomized controlled trials. JAMA, 292(23), 2874–2879.

    Article  PubMed  CAS  Google Scholar 

  4. Bristow, M. R., Saxon, L. A., Boehmer, J., Kruger, S., Kass, D. A., DeMarco, T. et al. (2004). Cardiac resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. New England Journal of Medicine, 350, 2140–2150.

    Article  PubMed  CAS  Google Scholar 

  5. Cleland, J. G. F., Daubert, J. C., Erdmann, E., Freemantle, N., Gras, D., Kappenberger, L., et al., the Cardiac Resynchronization—Heart Failure (CARE-HF) Study Investigators (2005). New England Journal of Medicine, 352, 1539–1549.

    Article  PubMed  CAS  Google Scholar 

  6. Horwich, T., Foster, E., De Marco, T., Tseng, Z., & Saxon, L. (2004). Effects of resynchronization therapy on cardiac function in pacemaker patients “upgraded” to biventricular devices. Journal of Cardiovascular Electrophysiology, 15(11), 1284–1289.

    Article  PubMed  Google Scholar 

  7. DeCara, J. M., Lang, R. M., Koch, R., Bala, R., Penzotti, J., & Spencer, K. T. (2002). The use of small personal ultrasound devices by internists without formal training in echocardiography. European Journal of Echocardiography, 4, 141–147.

    Article  Google Scholar 

  8. Kronzon, I. (2003). The hand-carried ultrasound revolution. Echocardiography, 20, 453–454.

    Article  PubMed  Google Scholar 

  9. Shen, W. K., Hammill, S. C., Hayes, D. L., Packer, D. L., Bailey, K. R., Ballard, D. J. et al. (1994). Long-term survival after pacemaker implantation for heart block in patients > or = 65 years. American Journal of Cardiology, 74, 560–564.

    Article  PubMed  CAS  Google Scholar 

  10. Lee, M. A., Dae, M. W., Langberg, J. J., Griffin, J. C., Chin, M. C., Finkbeiner, W. E. et al. (1994). Effects of long-term right ventricular apical pacing on left ventricular perfusion, innervation, function and histology. Journal of the American College of Cardiology, 24(1), 225–232.

    Article  PubMed  CAS  Google Scholar 

  11. Andersen, H. R., Nielsen, J. C., Thomsen, P. E., Thuesen, L., Mortensen, P. T., Vesterlund, T. et al. (1997). Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome. Lancet, 350(9086), 1210–1216.

    Article  Google Scholar 

  12. Thackray, S. D. R., Witte, K. K. A., Nikitin, N. P., Clark, A. L., Kaye, G. C., & Cleland, J. G. F. (2003). The prevalence of heart failure and asymptomatic left ventricular systolic dysfunction in a typical regional pacemaker population. European Heart Journal, 24, 1143–1152.

    Article  PubMed  Google Scholar 

  13. Decara, J. M., Kirkpatrick, J. N., Spencer, K. T., Ward, R. P., Kasza, K., Furlong, K. et al. (2005). Use of hand-carried ultrasound devices to augment the accuracy of medical student bedside cardiac diagnoses. Journal of the American Society of Echocardiography, 18, 257–263.

    Article  PubMed  Google Scholar 

  14. Duvall, W. L., Croft, L. B., & Goldman, M. E. (2003). Can hand-carried ultrasound devices be extended for use by the noncardiology medical community? Echocardiography, 20, 471–476.

    Article  PubMed  Google Scholar 

  15. Alexander, J. H., Peterson, E. D., Chen, A. Y., et al. (2001). Training and accuracy of non-cardiologists in simple use of point-of-care echo: A preliminary report from the Duke Limited Echo Assessment Project (LEAP). Thoraxcentre Journal, 13, 105–110.

    Google Scholar 

  16. Spencer, K. T., & Lang, R. M. (2002). Point of service echocardiography. ACC Current Journal Review, 11, 49–51.

    Article  Google Scholar 

  17. Kimura, B. J., Amundson, S. A., Willis, C. L., Gilpin, E. A., & DeMaria, A. N. (2002). Usefulness of a hand-hold ultrasound device for bedside examination of left ventricular function. American Journal of Cardiology, 90, 1038–1039.

    Article  PubMed  Google Scholar 

  18. Spencer, K. T., Fedson, S., Neithardt, G., Thomas, P., Lickerman, A., Radzienda, M. et al. (2002). Unsuspected clinically important cardiac findings detected with a small portable ultrasound device in patients admitted to a general medicine service (abstract). Circulation, 106, II-507.

    Google Scholar 

  19. Kobal, S. L., Tolstrup, K., Luo, H., Neuman, Y., Miyamoto, T., Mirocha, J., et al. (2004). Usefulness of a hand-carried cardiac ultrasound device to detect clinically significant valvular regurgitation in hospitalized patients. American Journal of Cardiology, 93, 1069–1072.

    Article  PubMed  Google Scholar 

  20. Bruce, C. J., Montgomery, S. C., Bailey, K. R., Tajik, J., & Seward, J. B. (2002). Utility of hand-carried ultrasound devices used by cardiologists with and without significant echocardiographic experience in the cardiology inpatient and outpatient settings. American Journal of Cardiology, 90, 1273–1275.

    Article  PubMed  Google Scholar 

  21. Rugolotto, M., Chang, C. P., Hu, B., Schnittger, I., & Liang, D. H. (2002). Clinical use of cardiac ultrasound performed with a hand-carried device in patients admitted for acute cardiac care. American Journal of Cardiology, 90, 1040–1042.

    Article  PubMed  Google Scholar 

  22. Kirkpatrick, J. N., Davis, A., Decara, J. M., Hong, A. E., Kurtz, P. L., Balasia, B. et al. (2004). Hand-carried cardiac ultrasound as a tool to screen for important cardiovascular disease in an underserved minority health care clinic. Journal of the American Society of Echocardiography, 17, 399–403.

    Article  PubMed  Google Scholar 

  23. Kirkpatrick, J. N., Belka, V., Furlong, K., Balasia, B., Jacobs, L. D., Corcoran, M. et al. (2005). Effectiveness of echocardiographic imaging by nurses to identify left ventricular systolic dysfunction in high-risk patients. American Journal of Cardiology, 95, 1271–1272.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to James N. Kirkpatrick.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ghani, S.N., Kirkpatrick, J.N., Spencer, K.T. et al. Rapid assessment of left ventricular systolic function in a pacemaker clinic using a hand-carried ultrasound device. J Interv Card Electrophysiol 16, 39–43 (2006). https://doi.org/10.1007/s10840-006-9011-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-006-9011-y

Keywords

Navigation