In situations in which it may be impossible and/or unethical to evaluate pulse oximetry in humans, an in vitro model with circulating blood may be a necessity. The main objective was to develop such an in vitro model and, in this model, validate the pulse oximetry technique at various haematocrit levels. The pulsating character of arterial blood flow in a tubing system was simulated by using a specially constructed pressure-regulated roller pump. The tubing system was designed to minimise damage to red blood cells. The pulse oximeter readings (SpO2) were compared with oxygen saturation analyses by a haemoximeter (SaO2). The pulse oximetry readings were recorded at various haematocrit levels and during haemolysis in the SaO2 range 60–100 per cent. At a haematocrit level of 41–44 per cent, there was no correlation between SaO2 and SpO2 readings. After diluting the blood with normal saline to a haematocrit of 10–11 per cent, a good correlation between SaO2 and SpO2 was found. Following haemolysis, the agreement between SaO2 and SpO2 was further improved. Using the developed in vitro model, the results indicate that the accuracy of a pulse oximeter may be dependent on the haematocrit level.
Similar content being viewed by others
References
Altman, D. G. andBland, J. M. (1983) Measurement in medicine: the analysis of method comparison studies.Statistician,32, 307–317.
Anderson, N. M. andSekelj, P. (1967) Light-absorbing and scattering properties of non-haemolysed blood.Phys. Med. Biol.,12, 173–184.
Barker, S. J. andTremper, K. K. (1987) The effect of carbon monoxide inhalation on pulse oximetry and transcutaneous PO2.Anesthesiol.,66, 677–679.
Bland, J. M. andAltman, D. G. (1986) Statistical methods for assessing agreement between two methods of clinical measurement.Lancet,i, 307–310.
Borgström, P., Clementz, L.-Å. andGrände, P.-O. (1981) A servo-controlled roller pump for constant flow or constant pressure blood perfusion under normal pulsatile or nonpulsatile conditions.Acta Physiol. Scand.,112, 437–442.
Brooks, T. D., Paulus, D. A. andWinkle, W. E. (1984) Intrared heat lamps interfere with pulse oximeters (letter).Anesthesiol.,61, 630.
Chien, S. (1975) Biophysical behavior of red cells in suspensions. InThe red blood cell.Surgenor, D. M. N. (Ed.) New York, Academic Press, Vol. 2, 1031–1133.
Coté, C. J., Goldstein, E. A., Coté, M. A., Hoaglin, D. C. andRyan, J. F. (1988) A single blind study of pulse oximetry in children.Anesthesiol.,68, 184–188.
de Kock, J. P. andTarassenko, L. (1991)In vitro investigation of the factors affecting pulse oximetry.J. Biomed. Eng.,13, 61–66.
Eisele, J. H. andDowns, D. (1987) Ambient light affects pulse oximeters.Anesthesiol.,67, 864–865.
Eisenkraft, J. B. (1988) Pulse oximeter desaturation due to methemoglobinemia. —Ibid.,68, 279–282.
Fairley, B. (1989) Changing perspectives in monitoring oxygenation. —Ibid.,70, 2–4.
Kelleher, J. F. (1989) Pulse oximetry.J. Clin. Monit.,5, 37–62.
Kelman, G. R. andNunn, J. F. (1966) Clinical recognition of hypoxaemia under fluorescent lamps.Lanceti, 1400–1403.
Klose, H. J., Volger, E., Brechtelsbauer, H., Heinich, L. andSchmid-Schönbein, H. (1972) Microrheology and light transmission of blood. I and II.Pflügers Arch.,333, 126–155.
Lee, S. E., Tremper, K. K. andBarker, S. J. (1988) Effects of anemia on pulse oximetry and continuous mixed venous hemoglobin saturation monitoring in dogs.Anesthesiol.,75, 118–122.
Lindberg, L.-G. (1991) Photoplethysmography—methodological studies and applications. Biomedical Engineering dissertation 262, Linkö** University, Sweden.
Lipowsky, H. H. andZweifach, B. W. (1977) Methods for simultaneous measurement of pressure differentials and flow in single unbranched vessels of the microcirculation for rheological studies.Microvasc. Res.,14, 345–361.
Mendelson, Y. andKent, J. C. (1989) An in vitro tissue model for evaluating the effect of carboxyhemoglobin concentration on pulse oximetry.IEEE Trans.,BME-36, 625–627.
Sakamoto, K. andKanai, H. (1979) Electrical characteristics of flowing blood. —Ibid.,BME-26, 686–695.
Scheller, M. S., Unger, R. J. andKelner, M. J. (1986) Effects of intravenously administered dyes on pulse oximetry readings.Anesthesiol.,65, 550–552.
Schmid-Schönbein, H. (1981) Interaction of vasomotor and blood rheology in haemodynamics. InClinical aspects of blood viscosity and cell deformability.Lowe, G. D. O., Barbenel, J. C. andForbes, C. D. (Eds.), Springer Verlag, New York, 56–57.
Sendak, M., Harris, A. andDonham, R. (1988) Accuracy of pulse oximetry during arterial oxyhemoglobin desaturation in dogs.Anesthesiol.,68, 111–114.
Severinghaus, J. W. andKoh, S. O. (1990) Effect of anemia on pulse oximeter accuracy at low saturation.J. Clin. Monit.,6, 85–88.
Shepherd, A. P., Riedel, G. L., Kiel, J. W., Haumschid, D. J. andMaxwell, L. C. (1987) Evaluation of an infrared laser-Doppler flowmeter.Am. J. Physiol.,252, G832-C839.
Shimada, Y., Yoshiya, I., Oka, N. andHamaguri, K. (1984) Effects of multiple scattering and peripheral circulation on arterial oxygen saturation measured with a pulse-type oximeter.Med. & Biol. Eng. & Comput.,22, 475–478.
Siegel, M. N. andGravenstein, N. (1987) Preventing ambient light from affecting pulse oximetry.Anesthesiol.,67, 280
Steinke, J. M. andShepherd, A. P. (1986) Role of light scattering in whole blood oximetry.IEEE Trans.,BME-33, 294–301.
Taylor, M. B. andWhitwam, J. G. (1988) The accuracy of pulse oximeters.Anaesthesia,43, 229–232.
Tremper, K. andBarker, S. (1989). Pulse oximetry.Anesthesiol.,70, 98–108.
Vegfors, M., Trygvason, B., Sjöberg, F. andLennmarken, C. (1990) Assessment of peripheral blood flow using a pulse oximeter.J. Clin. Monit.,6, 1–4.
Vegfors, M. andLennmarken, C. (1991) Carboxyhaemo-globinaemia and pulse oximetry.Br. J. Anaesth.,66, 625–626.
Vegfors, M., Lindberg, L.-G. Lennmarken, C. andÖberg, P. Å. (1991a) An in vitro model for evaluation of pulse oximetry. Proc. Optical Methods for Tumor Treat. and Early Diag.: Mechanisms and Techniques, SPIE Conf. Proc. 1426, 79–83.
Vegfors, M., Sjöberg, F., Lindberg, L.-G., Gustafsson, U. andLennmarken, C. (1991b) Basic studies of pulse oximetry in a rabbit model.Acta Anaesthesiol. Scand.,35, 596–599.
Vegfors, M., Lindberg, L.-G. andLennmarken, C. (1992) The influence of changes in blood flow on the accuracy of pulse oximetry in humans. —Ibid.,36, 346–349.
Yelderman, M. andNew, W. (1983) Evaluation of pulse oximetry.Anesthesiol.,59, 349–352.
Zdrojkowski, R. andPisharoty, J. N. R. (1970) Optical transmission and reflection by blood.IEEE Trans.,BME-17, 122–128.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Vegfors, M., Lindberg, L.G., Öberg, P.Å. et al. Accuracy of pulse oximetry at various haematocrits and during haemolysis in anin vitro model. Med. Biol. Eng. Comput. 31, 135–141 (1993). https://doi.org/10.1007/BF02446671
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02446671