Abstract
This study reports the method for measuring 3H and 14C in sludge samples from the National Physical Laboratory (NPL) proficiency program (Environmental Radioactivity Proficiency Test Exercise 2012). The samples were combusted by a high-temperature furnace and trapped beta-emitting nuclides, especially 3H and 14C, were measured by LSC. The validation of method was checked by using initial calibration verification (ICV) samples and laboratory control sample (LCS). The results were in agreement with assigned radioactivity concentration for 3H and 14C. Thus, reliable and accurate method for 3H and 14C was developed using high-temperature tube furnace and LSC.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10967-014-3425-2/MediaObjects/10967_2014_3425_Fig1_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10967-014-3425-2/MediaObjects/10967_2014_3425_Fig2_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10967-014-3425-2/MediaObjects/10967_2014_3425_Fig3_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10967-014-3425-2/MediaObjects/10967_2014_3425_Fig4_HTML.gif)
Similar content being viewed by others
References
Hou X (2005) Appl Radiat Isot 62:871–882
Bruce DL, Yechieli Y, Zilberbrand M, Kaufman A, Friedman GM (2007) J Hydrol 343:56–70
Boyer C, Vichot L, Fromm M, Losset Y, Tatin-Froux F, Guétat P, Badot PM (2009) Environ Exp Bot 67:34–51
Worth LBC, Pearce RJH, Bruce J, Banks J, Scales S (2005) Fusion Sci Technol 48:370–373
Soreefan AM, De Vol TA (2009) J Radioanal Nucl Chem 282:517–521
Hou X (2007) J Radioanal Nucl Chem 273:43–48
Caron F, Sutton J, Benz ML, Haas MK (2000) Analyst 125:59–64
Kutschera W (2013) Int J Mass Spectrom 349–350:203–218
Ognibene TJ, Bench G, Brown TA, Vogel JS (2004) Nucl Instrum Methods Phys Res B 223–224:12–15
Hou X, Roos P (2008) Anal Chim Acta 608:105–139
Warwick PE, Kim D, Croudace IW, Oh J (2010) Anal Chim Acta 676:93–102
Munakata K, Baba A, Nishikawa M, Penzhorn D (1998) J Nucl Sci Technol 35:511–514
Currie LA (1968) Anal Chem 40:586–593
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Lim, J.M., Kang, M.J., Chung, K.H. et al. Application of a high-temperature tube furnace and liquid scintillation counter for radioactivity determination of 14C and 3H in solid waste samples. J Radioanal Nucl Chem 303, 1111–1115 (2015). https://doi.org/10.1007/s10967-014-3425-2
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-014-3425-2