Abstract
Ideally, the flow delivery of an infusion system is proportional only to the rate of mechanical actuation of the syringe pump plunger. However, in the real world, overall infusion system compliance may be affected by components such as an extension of tubing lines, or different sizes of syringes. With higher compliance, there may be greater chances of flow irregularity. In this experimental study, we investigated the effects of lengths of infusion lines and syringe sizes on the compliance of syringe pumps with low flow rate (2 ml h−1). In the first experiment, infusion system compliance was measured in various settings by occlusion release. As the infusion tubing length and size of the syringe increased, the time to reach each pressure was delayed and the infusion system compliance increased. The contributions to system compliance from syringes were significantly greater compared to those of extended infusion lines. In the occlusion alarm experiment, the occlusion alarm could be delayed by 69.76 ± 3.98 min for the 50-ml syringe with a 560 cm infusion line set-up. In conclusion, the compliance of a syringe pump system increases as the loaded syringe size, or the length of the infusion tubing increases. The occlusion alarm may be much delayed and not useful in highly compliant systems with respect to the potential occlusion of the infusion system, so more attention is required when using a highly compliant infusion system.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10877-023-00973-y/MediaObjects/10877_2023_973_Fig1_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10877-023-00973-y/MediaObjects/10877_2023_973_Fig2_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10877-023-00973-y/MediaObjects/10877_2023_973_Fig3_HTML.png)
Similar content being viewed by others
References
Schulze KF, Graff M, Schimmel MS, Schenkman A, Rohan P. Physiologic oscillations produced by an infusion pump. J Pediatr. 1983;103:796–8. https://doi.org/10.1016/s0022-3476(83)80489-2.
Klem SA, Farrington JM, Leff RD. Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion. Crit Care Med. 1993;21:1213–7. https://doi.org/10.1097/00003246-199308000-00023.
Capes DF, Dunster KR, Sunderland VB, McMillan D, Colditz PB, McDonald C. Fluctuations in syringe-pump infusions: association with blood pressure variations in infants. Am J Health Syst Pharm. 1995;52:1646–53. https://doi.org/10.1093/ajhp/52.15.1646.
Thomas A, Galvin I. Patient safety incidents associated with equipment in critical care: a review of reports to the UK National Patient Safety Agency. Anaesthesia. 2008;63:1193–7. https://doi.org/10.1111/j.1365-2044.2008.05607.x.
Farbood A, Kazemi AP, Akbari K. Unintended purge during the start-up process of a syringe pump: report of a case presented with vascular collapse. J Clin Anesth. 2010;22:625–6. https://doi.org/10.1016/j.jclinane.2009.09.015.
Neff T, Fischer J, Fehr S, Baenziger O, Weiss M. Start-up delays of infusion syringe pumps. Paediatr Anaesth. 2001;11:561–5. https://doi.org/10.1046/j.1460-9592.2001.00730.x.
Neff SB, Neff TA, Gerber S, Weiss MM. Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps. Eur J Anaesthesiol. 2007;24:602–8. https://doi.org/10.1017/S0265021506002328.
Sarraf E, Mandel JE. Time-delay when updating infusion rates in the Graseby 3400 pump results in reduced drug delivery. Anesth Analg. 2014;118:145–50. https://doi.org/10.1213/01.ane.0000438349.21617.2f.
Murphy RS, Wilcox SJ. The link between intravenous multiple pump flow errors and infusion system mechanical compliance. Anesth Analg. 2010;110:1297–302. https://doi.org/10.1213/ANE.0b013e3181d56758.
Freckmann G, Kamecke U, Waldenmaier D, Haug C, Ziegler R. Occlusion detection time in insulin pumps at two different basal rates. J Diabetes Sci Technol. 2018;12:608–13. https://doi.org/10.1177/1932296817750404.
Neff T, Fischer J, Schulz G, Baenziger O, Weiss M. Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensiv Care Med. 2001;27:287–91. https://doi.org/10.1007/s001340000784.
Igarashi H, Obata Y, Nakajima Y, Katoh T, Morita K, Sato S. Syringe pump displacement alters line internal pressure and flow. Can J Anesth. 2005;52:685–91. https://doi.org/10.1007/BF03016554.
Hong KY, Kim YY, Yoo SY, Lee JH, Kim DK, Min JJ. Simulation study on flow rate accuracy of infusion pumps in vibration conditions during emergency patient transport. J Clin Monit Comput. 2020. https://doi.org/10.1007/s10877-020-00588-7.
Oh EJ, Hong KY, Lee JH, Kim DK, Cho J, Min JJ. Simulation analysis of flow rate variability during microinfusions: the effect of vertical displacement and multidrug infusion in conventional infusion pumps versus new cylinder-type infusion pumps. Anesth Analg. 2021. https://doi.org/10.1213/ANE.0000000000005736.
Coley SC, Shaw PK, Leff RD. The performance of three portable infusion-pump devices set to deliver 2 mL/hr. Am J Health Syst Pharm. 1997;54:1277–80. https://doi.org/10.1093/ajhp/54.11.1277.
Weiss M, Hug MI, Neff T, Fischer J. Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anaesth. 2000;47:1031–5. https://doi.org/10.1007/BF03024878.
Weiss M, Banziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensiv Care Med. 2000;26:776–9. https://doi.org/10.1007/s001340051246.
Weiss M, Neff T, Gerber A, Fischer J. Impact of infusion line compliance on syringe pump performance. Paediatr Anaesth. 2000;10:595–9. https://doi.org/10.1111/j.1460-9592.2000.566ab.x.
Schmidt N, Saez C, Seri I, Maturana A. Impact of syringe size on the performance of infusion pumps at low flow rates. Pediatr Crit Care Med. 2010;11:282–6. https://doi.org/10.1097/PCC.0b013e3181c31848.
Weiss M, Fischer J, Neff T, Baenziger O. The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia. 2000;55:1094–8. https://doi.org/10.1046/j.1365-2044.2000.01550.x.
Snijder RA, Konings MK, Lucas P, Egberts TC, Timmerman AD. Flow variability and its physical causes in infusion technology: a systematic review of in vitro measurement and modeling studies. Biomed Eng. 2015;60:277–300. https://doi.org/10.1515/bmt-2014-0148.
Chu DK, Akl EA, Duda S, Solo K, Yaacoub S, Schünemann HJ, Chu DK, Akl EA, El-harakeh A, Bognanni A, Lotfi T, Loeb M, Hajizadeh A, Bak A, Izcovich A, Cuello-Garcia CA, Chen C, Harris DJ, Borowiack E, Chamseddine F, Schünemann F, Morgano GP, Muti Schünemann GEU, Chen G, Zhao H, Neumann I, Chan J, Khabsa J, Hneiny L, Harrison L, Smith M, Rizk N, Giorgi Rossi P, AbiHanna P, El-khoury R, Stalteri R, Baldeh T, Piggott T, Zhang Y, Saad Z, Khamis A, Reinap M, Duda S, Solo K, Yaacoub S, Schünemann HJ. Physical distancing, face masks, and eye protection to prevent person-to-person transmission of SARS-CoV-2 and COVID-19: a systematic review and meta-analysis. Lancet. 2020;395:1973–87. https://doi.org/10.1016/s0140-6736(20)31142-9.
Clarkson DM, Barbosa R. Measurement of compliance of infusion device consumable elements using an analytical weighing balance. Med Eng Phys. 2014;36:1502–7. https://doi.org/10.1016/j.medengphy.2014.07.016.
Snijder RA, Knape JT, Egberts TC, Timmerman AM. Hypertensive crisis during norepinephrine syringe exchange: a case report. A&A Pract. 2017;8:178–81. https://doi.org/10.1213/XAA.0000000000000458.
Elli S, Mattiussi E, Bambi S, Tupputi S, San Fratello S, De Nunzio A, D’Auria S, Rona R, Fumagalli R, Lucchini A. Changing the syringe pump: a challenging procedure in critically ill patients. J Vasc Access. 2020;21:868–74. https://doi.org/10.1177/1129729820909024.
Min JJ, Kim DK, Hong KY, Choi JW, Choi KY. Comparison of operator workloads associated with the single-unit Anyfusion® pump and the changeover from a syringe pump to an infusion pump. J Korean Med Sci. 2019;34:10. https://doi.org/10.3346/jkms.2019.34.e314.
Donmez A, Araz C, Kayhan Z. Syringe pumps take too long to give occlusion alarm. Paediatr Anaesth. 2005;15:293–6. https://doi.org/10.1111/j.1460-9592.2005.01436.x.
Kariya T, Miyashita T, Sato H, Kawakami H, Goto T. Behavior of anesthesiology residents in a situation of intravenous route occlusion during syringe pump use in a simulated intensive care unit. J Patient Saf. 2019;15:290–2. https://doi.org/10.1097/PTS.0000000000000232.
van der Eijk AC, van Rens RM, Dankelman J, Smit BJ. A literature review on flow-rate variability in neonatal IV therapy. Pediatr Anesth. 2013;23:9–21. https://doi.org/10.1111/pan.12039.
Baeckert M, Batliner M, Grass B, Buehler PK, Daners MS, Meboldt M, Weiss M. Performance of modern syringe infusion pump assemblies at low infusion rates in the perioperative setting. Br J Anaesth. 2020;124:173–82. https://doi.org/10.1016/j.bja.2019.10.007.
Funding
The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.
Author information
Authors and Affiliations
Contributions
KYH, EKL, and JJM contributed to the study conception and design. KYH and EKL performed experiments. KYH, EKL, YYK, DCC and JJM analyzed data and interpreted results. KYH, EKL, and JJM drafted manuscript. KYH and JJM edited and revised the manuscript. All authors read and approved the final version of manuscript. JJM supervised the project.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
Ethical approval was waived by the Institutional Research Ethics Committee due to the in vitro experimental nature of this study.
Informed consent
No human subjects were included in this study, waiving the need for informed consent.
Research involving human participants and/or animals
This work did not to include research involving human participants and/or animals.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Hong, K.Y., Lee, E.K., Kim, Y. et al. Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate. J Clin Monit Comput 37, 1379–1386 (2023). https://doi.org/10.1007/s10877-023-00973-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10877-023-00973-y