Correction to: Analytical and Bioanalytical Chemistry

https://doi.org/10.1007/s00216-022-03875-1

Unfortunately, there were some errors in Tables 1 and 2 of this manuscript. The original article has been corrected; please find the correct information below:

Table 1 Precision and accuracy date for paraquat in plasma and lung tissue

Spiked paraquat (μg/L in plasma and μg/g in lung)

Mean measured paraquat (μg/L in plasma and μg/g in lung)

Precision (RSD, %)

Accuracy (%)

 

Plasma

  

Intra-day

1

0.99

8.3

98.9

2

1.74

6.2

86.9

50

49.04

5.2

98.1

400

420.43

11.1

105.1

Inter-day

1

0.98

16.5

98.0

2

1.84

11.5

91.9

50

53.03

8.1

106.1

400

383.52

12.1

95.9

 

Lung tissue

  

Intra-day

1

0.98

4.5

97.5

2

1.95

4.7

97.3

10

10.98

4.3

109.8

50

45.08

2.4

90.2

Inter-day

1

0.96

5.6

95.8

2

1.94

6.6

96.9

10

10.77

13.7

107.7

50

48.95

7.6

97.9

Table 2 Stability for paraquat in plasma

 

Spiked paraquat (μg/L or μg/g)

RSD (%)

Accuracy (%)

Plasma

Stability (RT, 4h)

2

6.4

94.8

50

3.9

102.0

400

8.6

96.8

2

6.2

87.7

Freeze-thaw stability

50

4.3

124.7

400

4.8

92.3

2

9.2

94.2

Long-term stability

50

3.3

114.1

400

4.0

100.5

Lung tissue

Stability (RT, 4h)

2

1.8

91.0

10

3.8

89.4

50

6.3

100.2

2

6.5

103.8

Freeze-thaw stability

10

4.9

124.7

50

7.0

104.3