Abstract
Analytical methods used for the isotope dilution inductively coupled plasma mass spectrometric (ID-ICP-MS) measurement of Cd at μg kg−1 and sub-μg kg−1 levels are described and applied to the certification of new dietary supplement, blood, and serum Standard Reference Materials (SRMs). The materials are: SRM 3240 Ephedra sinica Stapf Aerial Parts, SRM 3241 Ephedra sinica Stapf Native Extract, SRM 3243 Ephedra-Containing Solid Oral Dosage Form, SRM 3244 Ephedra-Containing Protein Powder, SRM 966 Toxic Metals in Bovine Blood, Level 1 (L1) and Level 2 (L2), and SRM 1598a Animal Serum. The concentration of Cd in the materials ranges from 120 μg kg−1 down to 0.03 μg kg−1. At these levels, the factors that most influence the accuracy of the ICP-MS data are the procedure blank and spectral and nonspectral interferences. Nonspectral interference, caused by the high concentration of dissolved solids in the matrices investigated, resulted in signal suppression. Matrix separation was used to enhance signal intensity and to reduce spectral interference for the accurate determination of Cd in SRM 1598a and SRM 3244. Chromatographic separation procedures using Chelex for SRM 1598a and anion exchange for SRM 3244 were optimized to achieve the desired separation characteristics without substantially increasing the procedure blank. Sensitivity for the determination of Cd in serum was additionally enhanced through the use of desolvation nebulization. We determined that separations were not required for the accurate ICP-MS determination of Cd in SRM 3240, SRM 3241, SRM 3243, and SRM 966 L2 under optimized analysis conditions. These samples were diluted to a minimum volume and introduced to the ICP-MS via low flow (40–100 μL/min) microconcentric nebulizers. SRM 966 L1 was also analyzed directly, but results were highly variable. The ID-ICP-MS sample preparation and ratio measurement protocols described here resulted in total expanded uncertainties of less than 1% for the determination of 90.85 μg kg−1 Cd in SRM 3240, and less than 10% total expanded uncertainty for the determination of 0.0468 μg kg−1 Cd in SRM 1598a.
Similar content being viewed by others
References
ATSDR (1999) Toxicological profile for cadmium. Agency for Toxic Substances and Disease Registry, Atlanta, GA, USA (see http://www.atsdr.cdc.gov/toxprofiles/tp5.pdf, last accessed 11th October 2006)
Wennberg M, Lundh T, Bergdahl IA, Hallmans G, Jansson JH, Stegmayr B, Custodio HM, Skerfving S (2006) Environ Res 100:330–338
Merrington G, Miller D, McLaughlin MJ, Keller MA (2001) Arch Environ Contam Toxicol 41:151–156
BurgatSacaze V, Craste L, Guere P (1996) Rev Med Vet–Toulouse 147:671–680
Zaho M, Wang J, Lu B, Lu H (2005) Rapid Commun Mass Sp 19:910–914
Sharpless KE, Anderson DL, Betz JM, Butler TA, Capar SG, Cheng J, Fraser CA, Gardner G, Gay ML, Howell DW, Ihara T, Khan MA, Lam JW, Long SE, McCooeye M, Mackey EA, Mindak WR, Mitvalsky S, Murphy KE, NguyenPho A, Phinney KW, Porter BJ, Roman M, Sander LC, Satterfield MB, Scriver C, Sturgeon R, Brown Thomas JM, Vocke RD Jr, Wise SA, Wood LJ, Yang L, Yen JH, Ziobro GC (2006) J AOAC Int (accepted)
Watters RL Jr, Eberhardt KR, Beary ES, Fassett JD (1997) Metrologia 34:87–96
Diemer J, Vogl J, Quetel CR, Linsinger T, Taylor PDP, Lamberty A, Pauwels J (2001) Fresenius J Anal Chem 370:492–498
Papadakis I, Quetel CR, Taylor PDP, De Bievre P (2000) Fresenius J Anal Chem 5:198–204
Beary ES, Paulsen PJ (1993) Anal Chem 65:1602–1608
Valles Mota JP, Fernández de la Campa MR, Garciá Alonso JI, Sanz-Medel A (1999) J Anal Atom Spectrom 14:113–120
Murphy KE, Beary ES, Rearick MS, Vocke RD (2000) Fresenius J Anal Chem 368:362–370
Kelly WR, Hotes SA (1988) J Res Natl Inst Stan 93:228–232
Fassett JD, Paulsen PJ (1989) Anal Chem 61:643A–649A
Nelson F, Rush RM, Kraus KA (1960) J Am Chem Soc 82:339–349
Yu LL, Vocke RD, Murphy KE, Beck CM (2001) Fresenius J Anal Chem 370:834–837
Russ GP III (1989) In: Date AR, Gray AL (eds) Applications of inductively coupled plasma mass spectrometry. Blackie, Glasgow, pp 90–140
Nonose N, Kubota M (2001) J Anal Atom Spectrom 16:560–566
Taylor BN, Kuyatt CE (1994) Guidelines for evaluating and expressing the uncertainty of NIST measurement results, 2nd edn (NIST Technical Note 1297). National Institute of Standards and Technology, Gaithersburg, MD (see http://physics.nist.gov/Pubs/guidelines/TN1297/tn1297s.pdf, last accessed 11th October 2006)
May W, Parris R, Beck C, Fassett J, Greenberg R, Guenther F, Kramer G, Wise S, Gills T, Colbert J, Gettings R, MacDonald B (2000) Definition of terms and modes used at NIST for value-assignment of reference materials for chemical measurements (NIST Special Publication 260-136). National Institute of Standards and Technology, Gaithersburg, MD (see https://srmors.nist.gov/referencelinks/references/SP260-136.pdf, last accessed 11th October 2006)
Disclaimer
Certain commercial equipment, instruments or materials are identified in this work to specify adequately the experimental procedure. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for this purpose.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Murphy, K.E., Long, S.E. & Vocke, R.D. On the certification of cadmium at trace and ultratrace levels in standard reference materials using ID ICP-MS. Anal Bioanal Chem 387, 2453–2461 (2007). https://doi.org/10.1007/s00216-006-0880-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-006-0880-9