Abstract
This is a report of our experience with a portable anaesthesia system that was developed for use under field conditions, when compressed gas supplies are limited. We first assembled and bench-tested a low-pressure plenum system, based upon the Farman entrainer. The entrainer required a low flow of compressed gas, O2 at 1–2 L⨴ min−1, and generated a low-pressure mixture of O2 and air which was directed through an Oxford miniature vaporizer, a non-return valve, and a wide-bore T-piece circuit. With this system we anaesthetized 24 patients with intermittent positive pressure ventilation (IPPV) and nine patients breathing spontaneously. During IPPV, the circuit resembled a T-piece and provided effective gas exchange with a FGF of 1.2 times minute ventilation. Inspiratory and expiratory valves were arranged so that the spontaneous mode was non-rebreathing, and FGF was adjusted to equal minute ventilation. The system was very economical, using 1–2 L · min−1 O2 and 20–25 ml · hr−1 liquid halothane to produce a FGF of 6–10 L min−1, an FlO2 of 0.33, and Flhal of 1–1.5 per cent. We have demonstrated that this is a versatile, safe, and economical system, compatible with the practice of modern inhalational anaesthesia under field conditions. It can be readily assembled from commercially available components.
Résumé
On reporte notre expérience avec un système d’anesthésie ambulant qui a été développé pour l’utilisation dans les conditions de champ de bataille quand la réserve de gaz comprimé est limitée. On a initialément assemblé et testé au laboratoire un systeme de soufflerie à basse pression basée sur le « Farman entrainer». Le souffleur a fonctionné à des flots bas de gaz comprimé O2 à 1–2 L·min−1, et genere un melange a basse pression d’oxygène et d’air el a été dirigé vers un vaporisateur miniature Oxford, une valve uni-directionnelle, et un circuit comprenant une piéce en T de grand calibre. Avec ce systeme, on a anesthésié 24 patients avec une ventilation intermittente à pression positive (IPPV) et les neuf patients raspiraient spontanément. Ourant l’ IPPV, le circuit ressemblait à la pièce en T et fournissait un échange gazeux efficace avec un FGF de 1.2 fois de ventilation à la minute. Les valves respiratoires et expiratoires ont été arrangées afin d’éviter la réinspiration lors de la ventilation spontanée et le FGF fut ajusté pour égaler la ventilation à la minute. Le systeme fut tres économique utilisant 1–2 L · min−1, O2 et 20–25 ml· heure−1 d’halothane liquide afin de produire un FGF de 6–10 L· min−1 et FlO2 de 0.33 et une Fl halothane de 1–1.5 pour cent. On a démontré que ce systeme versatile, sécuritaire et économique est compatible avec la pratique de l’anesthésie d’inhalation moderne dans les conditions de champs de bataille. Ce système pourrait être assemblé sur le champs à partir d’éléments disponibles commercialement.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Farman JV. Economical anaesthesia overseas: air entrainment device for use with draw-over vaporizers in children. Br Med J 1965; 1428.
Ezi-Ashi TI, Papworth DP, Nunn JF. Inhalational anaesthesia in develo** countries. Part I. The problem and a proposed solution. Anaesthesia 1983; 38: 729–5.
Farman JV. Anaesthesia and the EMO system. London: English University Press, 1973.
Merrifield AJ, Hill DW, Smith K. Performance of the Portablease and Fluoxair portable anaesthetic equipment. Br J Anaesth 1967; 39: 50–70.
Jowitt MD, Knight RJ. Anaesthesia during the Falklands Campaign. Anaesthesia 1983; 38: 776–83.
Parkhouse J. Clinical performance of the OMV vaporizer. Anaesthesia 1966; 21: 498–503.
Jensen JK. Halothankonzentrationen erreicht durch den “Oxford Miniature Vaporizer”. Der Anaesthetist 1967; 16: 54.
Wilson IH, Page RJE, Yeats MJH. The Oxford miniature vaporizer in paediatric anaesthesia. An experimental study. Anaesthesia 1988; 43: 700–2.
Carter JA, Baskett PJF, Simpson PJ. The “Permox” oxygen concentrator. Anaesthesia 1985; 40: 560–5.
Maltby JR, Rana NB, Amatya R, Shrestha BM. Anaesthesia training in Nepal. Can J Anaesth 1987; 34: 51–5.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Tweed, W.A., Amatya, R. & Spoerel, W.E. A low-pressure portable anaesthesia system for field use: clinical trials. Can J Anaesth 37, 928–931 (1990). https://doi.org/10.1007/BF03006639
Issue Date:
DOI: https://doi.org/10.1007/BF03006639