Abstract
Objectives
The objective of this study was to concurrently acquire an inductive k-space trajectory measure and corresponding imaging data by an MR scanner.
Materials and methods
1D gradient measures were obtained by digital integration, regularized using measured gradient coil currents and recorded individually by the scanner concurrently with raw MR data. Gradient measures were frequency modulated into an RF signal receivable by the scanner, yielding a k-space trajectory measure from the cumulative phase of the acquired data. Generation of the gradient measure and frequency modulation was performed by previously developed custom, versatile circuitry.
Results
For a normal echo planar imaging (EPI) sequence, the acquired k-space trajectory measure yielded slightly improved image quality compared to that obtained from using the scanner’s estimated eddy current-compensated k-space trajectory. For a spiral trajectory, the regularized inductive k-space trajectory measure lead to a 76% decrease in the root-mean-square error of the reconstructed image.
Discussion
While the proof-of-concept experiments show potential for further improvement, the feasibility of inductively measuring k-space trajectories and increasing the precision through regularization was demonstrated. The approach may offer an inexpensive method to acquire k-space trajectories concurrently with scanning.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig1_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig2_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig3_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig4_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig5_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig6_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10334-019-00770-2/MediaObjects/10334_2019_770_Fig7_HTML.png)
Similar content being viewed by others
References
Barmet C, Zanche ND, Pruessmann KP (2008) Spatiotemporal magnetic field monitoring for MR. Magn Reson Med 60(1):187–197
Dietrich BE, Brunner DO, Wilm BJ, Barmet C, Gross S, Kasper L, Haeberlin M, Schmid T, Vannesjo SJ, Pruessmann KP (2016) A field camera for MR sequence monitoring and system analysis. Magn Reson Med 75(4):1831–1840
Kasper L, Bollmann S, Vannesjo SJ, Gross S, Haeberlin M, Dietrich BE, Pruessmann KP (2015) Monitoring, analysis, and correction of magnetic field fluctuations in echo planar imaging time series. Magn Reson Med 74(2):396–409
Bruder H, Fischer H, Reinfelder HE, Schmitt F (1992) Image reconstruction for echo planar imaging with nonequidistant k-space sampling. Magn Reson Med 23(2):311–323
Delattre BMA, Heidemann RM, Crowe LA, Vallée JP, Hyacinthe JN (2010) Spiral demystified. Magn Reson Imaging 28(6):862–881
Duyn JH, Yang Y, Frank JA, van der Veen JW (1998) Simple correction method for k-space trajectory deviations in MRI. J Magn Reson 132(1):150–153
Mason GF, Harshbarger T, Hetherington HP, Zhang Y, Pohost GM, Twieg DB (1997) A method to measure arbitrary k-space trajectories for rapid MR imaging. Magn Reson Med 38(3):492–496
Onodera T, Matsui S, Sekihara K, Kohno H (1987) A method of measuring field-gradient modulation shapes. Application to high-speed NMR spectroscopic imaging. J Phys E Sci Inst 20(4):416–419
Addy NO, Wu HH, Nishimura DG (2012) Simple method for MR gradient system characterization and k-space trajectory estimation. Magn Reson Med 68(1):120–129
Campbell-Washburn AE, Xue H, Lederman RJ, Faranesh AZ, Hansen MS (2016) Real-time distortion correction of spiral and echo planar images using the gradient system impulse response function. Magn Reson Med 75(6):2278–2285
Vannesjo SJ, Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP (2013) Gradient system characterization by impulse response measurements with a dynamic field camera. Magn Reson Med 69(2):583–593
Brunner DO, Dietrich BE, Cavusoglu M, Wilm BJ, Schmid T, Gross S, Barmet C, Pruessmann KP (2016) Concurrent recording of RF pulses and gradient fields—comprehensive field monitoring for MRI. NMR Biomed 29(9):1162–1172
Vannesjo SJ, Dietrich BE, Pavan M, Brunner DO, Wilm BJ, Barmet C, Pruessmann KP (2014) Field camera measurements of gradient and shim impulse responses using frequency sweeps. Magn Reson Med 72(2):570–583
Barmet C, De Zanche N, Wilm BJ, Pruessmann KP (2009) A transmit/receive system for magnetic field monitoring of in vivo MRI. Magn Reson Med 62(1):269–276
Senaj V, Guillot G, Darrasse L (1998) Inductive measurement of magnetic field gradients for magnetic resonance imaging. Rev Sci Inst 69(6):2400–2405
D’Antona G, Lazzaroni M, Ottoboni R, Svelto C (2003) AC current-to-voltage transducer for industrial application. In: Instrumentation and measurement technology: 2003 IEEE conference, pp 1185–1190
Hanson LG, Lund TE, Hanson CG (2007) Encoding of electrophysiology and other signals in MR images. J Magn Reson Imaging 25(5):1059–1066
Pedersen JO, Hanson CG, Xue R, Hanson LG (2018) General purpose electronics for real-time processing and encoding of non-MR data in MR acquisitions. Concepts Magn Reson Part B Magn Reson Eng 48(2):e21385
Hanson LG (2018) Encoding of Non-MR signals in scanner recordings. http://drcmr.dk/MagstripeEncoding. Accessed 26 July 2019
Bernstein MA, King KF, Zhou XJ (2004) Handbook of MRI pulse sequences. Elsevier Academic Press, Burlington
Brodsky EK, Samsonov AA, Block WF (2009) Characterizing and correcting gradient errors in non-cartesian imaging: are gradient errors linear time-invariant (LTI)? Magn Reson Med 62(6):1466–1476
Spielman DM, Pauly JM (1995) Spiral imaging on a small-bore system at 4.7t. Magn Reson Med 34(4):580–585
Hiltunen J, Hari R, Jousmäki V, Müller K, Sepponen R, Joensuu R (2006) Quantification of mechanical vibration during diffusion tensor imaging at 3 T. NeuroImage 32(1):93–103
Greengard L, Lee JY (2004) Accelerating the Nonuniform Fast Fourier Transform. SIAM Rev 46(3):443–454
Pipe JG, Menon P (1999) Sampling density compensation in MRI: rationale and an iterative numerical solution. Magn Reson Med 41(1):179–186
De Zanche N, Barmet C, Nordmeyer-Massner Ja, Pruessmann KP (2008) NMR Probes for measuring magnetic fields and field dynamics in MR systems. Magn Reson Med 60(1):176–186
Wilm BJ, Barmet C, Pavan M, Pruessmann KP (2011) Higher order reconstruction for MRI in the presence of spatiotemporal field perturbations. Magn Reson Med 65(6):1690–1701
Andersen M, Hanson LG, Madsen KH, Wezel J, Boer V, Van Der Velden T, Van Osch MJ, Klomp D, Webb AG, Versluis MJ (2016) Measuring motion-induced B0-fluctuations in the brain using field probes. Magn Reson Med
Analog Devices; Inc (2016) AD9951 direct digital synthesize (data sheet. Rev, B)
Nevo E (2003) Method and apparatus to estimate location and orientation of objects during magnetic resonance imaging. US patent 6,516,213 B1, 4 Feb 2003
van Niekerk A, Meintjes E, van der Kouwe A (2019) A wireless radio frequency triggered acquisition device (WRAD) for self-synchronised measurements of the rate of change of the MRI gradient vector field for motion tracking. IEEE Trans Med Imaging 38(7):1610–1621
Acknowledgements
We acknowledge Dr. Zhentao Zuo for his kind support. We would also like to thank the reviewers for their many good suggestions for improving the paper.
Funding
This article is funded by Sino-Danish Center for Education and Research.
Author information
Authors and Affiliations
Contributions
Study conception and design were performed by all authors. Acquisition of data was performed by Pedersen and L. Hanson using hardware developed by C. Hanson. Analysis and interpretation of data were done by all authors. Drafting of manuscript was done by Pedersen and L. Hanson. Critical revision was done by all authors.
Corresponding author
Ethics declarations
Conflict of interest
After completion of the presented work, Pedersen has become an employee of Philips Healthcare.
Ethical approval
No human subjects or animals were scanned in this study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Pedersen, J.O., Hanson, C.G., Xue, R. et al. Inductive measurement and encoding of k-space trajectories in MR raw data. Magn Reson Mater Phy 32, 655–667 (2019). https://doi.org/10.1007/s10334-019-00770-2
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10334-019-00770-2