Abstract
This paper presents the direct comparison between measurements and Monte Carlo ion chamber simulations employing the EGSnrc Monte Carlo code.
Surrounding metal caps for a 1 cm3 ionization chamber (PTW30015) were varied both, in measurements and simulations. Absorbing caps made of iron, copper and aluminum alloy were manufactured each with wall thickness between 2 and 30 mm. As a radiation source, a Siemens Gammatron S 60Co machine was used. Normalized dose values were compared and changes to the cross-sections used in the Monte Carlo simulations investigated.
Results agreed within the range of the measurement uncertainties of 0.5%. Still, variations in photon cross-sections lead to significant changes in the results, highlighting the need of accurate input data for this type of investigation.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Toussaint, A., Wulff, J., Zink, K. (2009). Measurements and EGSnrc Monte Carlo Simulations for Thimble Ion Chambers with Various Metal Caps at 60Co Beams. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/3. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03902-7_131
Download citation
DOI: https://doi.org/10.1007/978-3-642-03902-7_131
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-03901-0
Online ISBN: 978-3-642-03902-7
eBook Packages: EngineeringEngineering (R0)