Log in

A Novel Electromechanical Intramedullary Nail System for Limb Lengthening

  • Original Article
  • Published:
Indian Journal of Orthopaedics Aims and scope Submit manuscript

Abstract

Background/Purpose of the Study

Many methods have been developed to treat leg length discrepancies. Extensible intramedullary nails are the most commonly used systems. However, complications such as excessive distraction, blockage of the nail, aditional surgery, uncontrolled lengthening and much pain occur during the use of these systems, and the desired success rate cannot be achieved. This study aims to develop a new extensible intramedullary system with two mechanisms (electronical and mechanical) for femur and tibia that treats leg length inequalities in a way that allows lengthening without complications as much as possible and does not require a second surgical intervention. It was planned to perform basic mechanical and cadaver functional tests of the new system, which will be designed and developed for this purpose.

Methods

The 3D design of the system has been completed with 3D computer software. A compact system has been developed that is mechanically activated by sudden axial loading and electronically activated by a controllable electric motor. Basic mechanical and functional tests of the new system have been performed within a cadaver.

Results

The rapid prototype of the system with electronical and mechanical units has been produced. As a result of the mechanical tests (axial loading), the nail was found to be resistant to compression forces. Its application to the cadaver and function tests was successfully performed.

Conclusions

We believe that the system we have developed will have advantages, such as working principle, ease of application, controlled lengthening, patient mobility compared to existing leg lengthening methods. The success of the system in practice will be evaluated by in-vivo animal experiments after more detailed mechanical experiments on cadavers. According to the results, it will be ready for human use by performing necessary restorations.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
EUR 32.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or Ebook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price includes VAT (France)

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10

Similar content being viewed by others

Data availability

The data that support the findings of this study are available from the corresponding author, [Bora UZUN], upon reasonable request.

References

  1. Götz, J., & Schellmann, W. D. (1975). Continuous lengthening of the femur with intramedullary stabilisation (author’s transl). Archiv fur Orthopadische und Unfall-Chirurgie [Internet], 82(4), 305–310.

    Article  PubMed  Google Scholar 

  2. Baumann, F., & Harms, J. (1977). Der Verlangerungsnagel (The extension nail A new method for lengthening of the femur and tibia). Archiv fur Orthopedische und Unfall-Chirurgie [Internet], 90(2), 139–146. https://doi.org/10.1007/BF00414987

    Article  CAS  Google Scholar 

  3. Hasler, C. C., & Krieg, A. H. (2012). Current concepts of leg lengthening. Journal of Children’s Orthopaedics, 6(2), 89–104.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Bliskunov, A. I. (1983). Intramedullary distraction of the femur (preliminary report). Ortopediia Travmatologiia i Protezirovanie, 10, 59–62.

    Google Scholar 

  5. Paley, D., Harris, M., Debiparshad, K., & Prince, D. (2014). Limb lengthening by implantable limb lengthening devices. Techniques in Orthopaedics [Internet], 29(2), 72–85. Available from: https://journals.lww.com/techortho/Fulltext/2014/06000/Limb_Lengthening_by_Implantable_Limb_Lengthening.5.aspx

  6. Guichet, J. M., & Casar, R. S. (1997). Mechanical characterization of a totally intramedullary gradual elongation nail. Clinical Orthopaedics and Related Research [Internet], 337(337), 281–290. Available from: https://pubmed.ncbi.nlm.nih.gov/9137201/

  7. Guichet, J. M., Deromedis, B., Donnan, L. E. O. T., Peretti, G., Lascombes, P., & Bado, F. (2003). Gradual femoral lengthening with the albizzia intramedullary nail. The Journal of Bone and Joint Surgery-American Volume., 85(5), 838–848.

    Article  PubMed  Google Scholar 

  8. Hankemeier, S., Gösling, T., Pape, H. C., Wiebking, U., & Krettek, C. (2005). Limb lengthening with the intramedullary skeletal kinetic distractor (ISKD). Operative Orthopädie und Traumatologie, 17(1), 79–101.

    Article  PubMed  Google Scholar 

  9. Baumgart, R., Betz, A., & Schweiberer, L. (1997). A fully implantable motorized intramedullary nail for limb lengthening and bone transport. Clinical Orthopaedics and Related Research [Internet], 343, 135–143. Available from: http://www.ncbi.nlm.nih.gov/pubmed/9345218

  10. Baumgart, R., Thaller, P., Hinterwimmer, S., Krammer, M., Hierl, T., & Mutschler, W. (2006). A fully implantable, programmable distraction nail (fitbone)—New perspectives for corrective and reconstructive limb surgery. Practice of intramedullary locked nails (pp. 189–198). Springer.

    Google Scholar 

  11. Betz, A., Baumgart, R., & Schweiberer, L. (1990). First fully implantable intramedullary system for callus distraction—intramedullary nail with programmable drive for leg lengthening and segment displacement. Principles and initial clinical results. Chirurg, 61(8), 605–609.

    CAS  PubMed  Google Scholar 

  12. Havitcioglu, H., Baran, Ö., Uzun, B., & Oflaz, H. (2007). Intramedullary fixation pin, Turkiye.

  13. Uzun, B. (2008). Improving of intramedullary nails used on bone defects and extremity lengthening and design of a new intramedullary nail. Dokuz Eylul University.

  14. Ertem, F. (2011). Development of lengthening plate system which can be used extra-skeletally in the body. Dokuz Eylul University.

  15. Aliç, T., Aliç, T., Güler, C., Çalbiyik, M., & Hassa, E. (2023). Health sciences medicine which of the three different intramedullary nail designs is superior in the treatment of femoral shaft fractures? Journal of Health Sciences and Medicine, 6(2), 467–475.

    Article  Google Scholar 

  16. García-Cimbrelo, E., de la Mano, A. C., García-Rey, E., Cordero, J., & Marti-Ciruelos, R. (2002). The intramedullary elongation nail for femoral lengthening. The Journal of Bone and Joint Surgery, 84(7), 971–977.

    Article  PubMed  Google Scholar 

  17. Guichet, J. M., Lascombes, P. F. M., Grammont, P. M., & Prévot, J. M. (1995). Gradual elongation intramedullary nail for femur (Albizzia). Results of the 52 first cases in 48 patients. Journal of Orthopaedic Science, 1995, 69.

    Google Scholar 

  18. Reynders, P. (2009). The runaway nail. Injury Extra, 40(9), 165–167.

    Article  Google Scholar 

  19. Cole, J. D., Justin, D., Kasparis, T., DeVlught, D., & Knobloch, C. (2001). The intramedullary skeletal kinetic distractor (ISKD): First clinical results of a new intramedullary nail for lengthening of the femur and tibia. Injury [Internet], 32, SD-129-SD−139.

    Article  Google Scholar 

  20. Kenawey, M., Krettek, C., Liodakis, E., Wiebking, U., & Hankemeier, S. (2011). Leg lengthening using intramedullay skeletal kinetic distractor: Results of 57 consecutive applications. Injury, 42(2), 150–155.

    Article  PubMed  Google Scholar 

  21. Kubiak, E. N., Strauss, E., Grant, A., Feldman, D., & Egol, K. A. (2007). Early complications encountered using a self-lengthening intramedullary nail for the correction of limb length inequality (pp. 52–57). Available from: https://jointdrs.org/full-text-pdf/497

  22. Mahboubian, S., Seah, M., Fragomen, A. T., & Rozbruch, S. R. (2012). Femoral lengthening with lengthening over a nail has fewer complications than intramedullary skeletal kinetic distraction. Clinical Orthopaedics and Related Research, 470(4), 1221–1231.

    Article  PubMed  Google Scholar 

  23. Hankemeier, S., Pape, H. C., Gosling, T., Hufner, T., Richter, M., & Krettek, C. (2004). Improved comfort in lower limb lengthening with the intramedullary skeletal kinetic distractor. Archives of Orthopaedic and Trauma Surgery [Internet], 124(2), 129–133. https://doi.org/10.1007/s00402-003-0625-6

    Article  PubMed  Google Scholar 

  24. Simpson, A. H. W. R., Shalaby, H., & Keenan, G. (2009). Femoral lengthening with the intramedullary skeletal kinetic distractor. Journal of Bone and Joint Surgery [Internet], 91(7), 955–961.

    CAS  Google Scholar 

  25. Accadbled, F., Pailhé, R., Cavaignac, E., & Sales de Gauzy, J. (2016). Bone lengthening using the Fitbone® motorized intramedullary nail: The first experience in France. Orthopaedics and Traumatology: Surgery and Research, 102(2), 217–222.

    CAS  Google Scholar 

  26. Horn, J., Grimsrud, Ø., Dagsgard, A. H., Huhnstock, S., & Steen, H. (2015). Femoral lengthening with a motorized intramedullary nail. Acta Orthopaedica, 86(2), 248–256.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Calder, P. R., McKay, J. E., Timms, A. J., Roskrow, T., Fugazzotto, S., Edel, P., & Goodier, W. D. (2019). Femoral lengthening using the Precice intramedullary limb-lengthening system: Outcome comparison following antegrade and retrograde nails. Bone and Joint Journal, 101-B(9), 1168–1176.

    Article  PubMed  Google Scholar 

  28. Calder, P. R., Wright, J., & Goodier, W. D. (2022). An update on the intramedullary implant in limb lengthening: A quinquennial review Part 1: The further influence of the intramedullary nail in limb lengthening. Injury [Internet], 53, S81–S87.

    Article  PubMed  Google Scholar 

  29. Frost, M. W., Rahbek, O., Traerup, J., Ceccotti, A. A., & Kold, S. (2020). Systematic review of complications with externally controlled motorized intramedullary bone lengthening nails (FITBONE and PRECICE) in 983 segments. Acta Orthopaedica, 92(1), 120–127.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Hosny, G. A. (2020). Limb lengthening history, evolution, complications and current concepts. Journal of Orthopaedics and Traumatology, 21(1), 1.

    Article  Google Scholar 

  31. Masci, G., Palmacci, O., Vitiello, R., Bonfiglio, N., Bocchi, M. B., Cipolloni, V., Maccauro, G., & Pola, E. (2021). Limb lengthening with PRECICE magnetic nail in pediatric patients: A systematic review. World Journal of Orthopedics [Internet], 12(8), 575–583.

    Article  PubMed  Google Scholar 

  32. Rentenberger, C., Haider, T., Zak, L., & Wozasek, G. E. (2021). Implant failure after externally magnetic-controlled nail lengthening. Injury, 52(11), 3350–3354.

    Article  PubMed  Google Scholar 

  33. Wiebking, U., Liodakis, E., Kenawey, M., & Krettek, C. (2016). Limb lengthening using the PRECICETM nail system: Complications and results. Archives of Trauma Research, 5(4), 1.

    Article  Google Scholar 

  34. Paley, D., Harris, M., Debiparshad, K., & Prince, D. (2014). Limb lengthening by implantable limb lengthening devices. Techniques in Orthopaedics [Internet] [cited 2022 November 22];29(2):72–85. Available from: https://journals.lww.com/techortho/Fulltext/2014/06000/Limb_Lengthening_by_Implantable_Limb_Lengthening.5.aspx

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bora Uzun.

Ethics declarations

Conflct of Interest

On behalf of all authors, the corresponding author states that there is no conflict of interest.

Ethical Standard Statement

This article does not contain any studies with human or animal subjects performed by the any of the authors.

Informed Consent

For this type of study, informed consent is not required.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Uzun, B., Havıtçıoğlu, H. A Novel Electromechanical Intramedullary Nail System for Limb Lengthening. JOIO 57, 1765–1776 (2023). https://doi.org/10.1007/s43465-023-00966-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s43465-023-00966-8

Keywords

Navigation