Radiation dose evaluations during radiological contrast studies in patients with morbid obesity

Valutazioni dosimetriche in esami contrastografici eseguiti su pazienti affetti da obesità patologica

  • Radiobiology/Radiobiologia
  • Published:
La radiologia medica Aims and scope Submit manuscript

Abstract

Purpose

The purpose of this study was to evaluate the radiation dose to patients during radiological contrast studies performed after vertical banded gastroplasty (VBG) or Roux-en-Y gastric bypass (RYGBP) surgery in patients with morbid obesity.

Materials and methods

Dose evaluations were performed on a sample of 39 patients (32 women and 7 men) with a mean weight of 117 kg (range 68–175 kg) and a mean body mass index (BMI) of 43.7 (range 22.2–54.9). Between the second and seventh postoperative day, patients underwent radiological follow-up after oral administration of approximately 70 ml of water-soluble iodinated contrast material (Gastrografin) and images acquired in anteroposterior, right and left oblique projections with the patient upright and then supine. Exposure conditions, dose-area product (DAP) and entrance skin dose (ESD) were recorded for each procedure. On the basis of these data, the effective dose (ED) was calculated using simulation software based on the Monte Carlo method for determining the absorbed dose to organs. To assess the optimal exposure conditions and the dose contributions of fluoroscopy and radiography, the effective dose rates were also evaluated using Plexiglas phantoms of different thickness to simulate different patient sizes.

Results

The phantom measurements showed a fourfold dose increase when passing from normal-sized patients to obese patients. Mean DAP value obtained from in-vivo measurements was 70 Gy cm2 (range 17–147 Gy cm2), and mean effective dose was 21 mSv (range 5–45 mSv).

Conclusions

When performing radiological contrast studies in patients with morbid obesity, every possible precaution should be taken to minimise patient dose. Special care should be taken to evaluate justification of the radiological procedure.

Riassunto

Obiettivo

Scopo del lavoro è stato quello di valutare la dose al paziente nel corso di esami contrastografici eseguiti in pazienti affetti da obesità patologica sottoposti ad interventi chirurgici di gastroplastica o bypass gastrico.

Materiali e metodi

Le valutazioni sono state effettuate per 39 esami, su di un campione di pazienti costituito da 32 donne e 7 uomini, con peso medio di 117 kg (intervallo 68–175 kg) e con indice di massa corporeo medio di 43,7 (intervallo 22,2–54,9). I pazienti operati nell’arco di tempo della durata dello studio sono stati sottoposti a controllo radiologico post-operatorio tra la II e la VII giornata mediante somministrazione per os di circa 70 ml di MdC iodato idrosolubile (Gastrografin) e acquisizione di radiogrammi eseguiti dapprima in ortostatismo in proiezione AP, obliqua destra e sinistra e successivamente a paziente supino. Per l’intera durata di ciascuna procedura sono state registrate le condizioni di esposizione e i valori di prodotto dose per area (DAP) e di dose di ingresso al paziente (ESD). Sulla base dei dati rilevati è stato effettuato un calcolo della dose efficace (ED) al paziente utilizzando un programma di calcolo basato sul metodo Monte Carlo per la stima della dose assorbita agli organi. Al fine di valutare le condizioni ottimali di esposizione e i contributi alla dose totale legati alla scopia e alla grafia, sono state inoltre effettuate misure di dose con fantocci di plexiglass di diversi spessori per simulare i pazienti di diversa corporatura.

Risultati

Le misurazioni su fantoccio hanno dimostrato che il passaggio da spessori analoghi a quelli dei pazienti normotipo a quelli dei pazienti oggetto dello studio può comportare un aumento di dose di un fattore quattro. Il valore medio di prodotto dose per area ottenuto nelle misurazioni in vivo è stato di 70 Gy cm2 (intervallo 17–147 Gy cm2), mentre il valore medio di dose efficace è stato di 21 mSv (intervallo 5–45 mSv).

Conclusioni

Nell’esecuzione di esami contrastografici in pazienti obesi occorre adottare tutte le precauzioni possibili per ridurre al minimo la dose al paziente. Particolare attenzione deve essere posta alla giustificazione dell’esame.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
EUR 32.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or Ebook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price includes VAT (United Kingdom)

Instant access to the full article PDF.

Similar content being viewed by others

References/Bibliografia

  1. Mason EE, Printen KJ, Blommers TJ et al (1980) Gastric bypass in morbid obesity. Am J Clin Nutr 33:395–405

    PubMed  CAS  Google Scholar 

  2. European Association for Endoscopic Surgery (E.A.E.S.) (2005) Evidencebased guidelines of the European Association for Endoscopic Surgery (E.A.E.S.). Surg Endosc 19:200–221

    Article  Google Scholar 

  3. Toppino M, Cesarani F, Comba A et al (2001) The role of early radiological studies after gastric bariatric surgery. Obes Surg 11:447–454

    Article  PubMed  CAS  Google Scholar 

  4. Martin CJ (2004) A review of factors affecting patient doses for barium enemas and meals. Brit J Radiol 77:864–868

    Article  PubMed  CAS  Google Scholar 

  5. Bryk SG, Censullo ML, Wagner LK et al (2006) Endovascular and interventional procedures in obese patients: a review of procedural technique modifications and radiation management. J Vasc Interv Radiol 17:27–33

    Article  PubMed  Google Scholar 

  6. ICRP (2001) Avoidance of radiation injuries from medical interventional procedures. ICR Publication 85

  7. Tapiovaarra M, Lakkisto M, Servomaa A (1997) PCXMC. A PC based Monte Carlo program for calculating patient doses in medical x-ray examinations. Helsinki: Finnish Centre for Radiation and Nuclear Safety. STUKA 139

    Google Scholar 

  8. Geleijns J, Broerse JJ, Chandie Shaw MP et al (1998) A comparison of patient dose for examinations of the upper gastrointestinal tract at 11 conventional and digital X-ray units in The Netherlands. Brit J Radiol 71:745–753

    PubMed  CAS  Google Scholar 

  9. Carroll EM, Brennan PC (2003) Radiation doses for barium enema and barium meal examinations in Ireland: potential diagnostic reference levels. Brit J Radiol 76:393–397

    Article  PubMed  CAS  Google Scholar 

  10. Delichas MG, Hatziioannou K, Papanastassiou E et al (2004) Radiation doses to patients undergoing barium meal and barium enema examinations. Rad Prot Dosim 109:243–247

    Article  CAS  Google Scholar 

  11. Morrell RE, Rogers AT, Jobling JC et al (2004) Barium Enema: use of increased copper filtration to optimize dose and image quality. Brit J Radiol 77:116–122

    Article  PubMed  CAS  Google Scholar 

  12. Moro L, Cazzani C, Tomarchio O et al (2007) Patient dose during radiological examination in the follow up of bariatric surgery. Rad Prot Dosimetry 123:113–117

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to O. Rampado.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Rampado, O., Luberto, L., Faletti, R. et al. Radiation dose evaluations during radiological contrast studies in patients with morbid obesity. Radiol med 113, 1229–1240 (2008). https://doi.org/10.1007/s11547-008-0335-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11547-008-0335-9

Keywords

Parole chiave

Navigation