Abstract
Background
In recent years, the use of fibrin glue has become an established practice in several areas of surgical treatment. For example, fibrin glue is used increasingly as an alternative method for mesh fixation in hernia surgery, significantly hel** to reduce the incidence of chronic pain. The experiments in this study were aimed at elucidating the extent to which tack- or suture-based permanent fixation can be replaced by fixation with fibrin glue for laparoscopic intraperitoneal repair of abdominal wall hernias.
Methods
In an initial series of experiments conducted with a biomechanical model, the strength of the fibrin glue for fixation of lightweight mesh (TiMesh light) to muscle tissue was compared with its strength of fixation to the peritoneum. In a second series of experiments, mesh was laparoscopically implanted in an established porcine model. Fibrin glue was used for mesh fixation in six animals. Laparoscopic exploration and explantation of the meshes were conducted after 4 months. Planimetric analysis was performed to investigate adhesion and shrinkage of the mesh surface.
Results
The strength of fibrin glue for fixation of mesh to the peritoneum was significantly less than for its fixation to muscle tissue (11.86 N vs. 47.88 N; p = 0.001). Three of the implanted meshes were not completely integrated, and two of these were dislocated. On the average, adhesions were seen on 16% of the mesh surfaces. The mesh shrinkage rate was 24.2%.
Conclusion
Mesh fixation alone to the undamaged peritoneum in the intraperitoneal region cannot be recommended because of the risk for dislocation. Additional fixation using sutures, tacks, or both is needed until the mesh material is completely integrated.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00464-009-0509-0/MediaObjects/464_2009_509_Fig1_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00464-009-0509-0/MediaObjects/464_2009_509_Fig2_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00464-009-0509-0/MediaObjects/464_2009_509_Fig3_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00464-009-0509-0/MediaObjects/464_2009_509_Fig4_HTML.jpg)
Similar content being viewed by others
References
Cuschieri A (2001) Tissue adhesives in endosurgery. Semin Laparosc Surg 8:63–68
Lee MG, Jones D (2005) Applications of fibrin sealant in surgery. Surg Innov 12:203–213
Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233:18–25
Katkhouda N (2004) A new technique for laparoscopic hernia repair using fibrin sealant. Surg Technol Int 12:120–126
Olmi S, Addis A, Domeneghini C, Scaini A, Croce E (2007) Experimental comparison of type of tissucol dilution and composite mesh (Parietex) for laparoscopic repair of groin and abdominal hernia: observational study conducted in a university laboratory. Hernia 11:211–215
ten Hallers EJ, Jansen JA, Marres HA, Rakhorst G, Verkerke GJ (2007) Histological assessment of titanium and polypropylene fiber mesh implantation with and without fibrin tissue glue. J Biomed Mater Res A 80:372–380
Topart P, Vandenbroucke F, Lozach P (2005) Tisseel versus tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis. Surg Endosc 19:724–727
Schwab R, Willms A, Kröger A, Becker HP (2006) Less chronic pain following mesh fixation using a fibrin sealant in TEP inguinal hernia repair. Hernia 10:272–277
Lovisetto F, Zonta S, Rota E, Mazzilli M, Bardone M, Bottero L, Faillance G, Longoni M (2007) Use of human fibrin glue (Tissucol) versus staples for mesh fixation in laparoscopic transabdominal preperitoneal hernioplasty. Ann Surg 245:222–231
Olmi S, Scaini A, Erba L, Guaglio M, Croce E (2007) Quantification of pain in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernioplasty idendifies marked differences between prosthesis fixation systems. Surgery 142:40–46
Nienhuis S, Staal E, Strobbe L, Rosman C, Groenewoud H, Bleichrodt R (2007) Chronic pain after mesh repair of inguinal hernia: a systematic review. Am J Surg 194:394–400
Morris-Stiff G, Hughes LE (1998) The outcomes of nonabsorbable mesh placed within the abdominal cavity: literature review and clinical experience. J Am Coll Surg 186:352–357
Rudmik LR, Schieman C, Dixon E, Debru E (2006) Laparoscopic incisional hernia repair: a review of the literature. Hernia 10:110–119
Heniford BT, Park A, Ramshaw BJ, Voeller G (2003) Laparoscopic repair of ventral hernias: nine years experience with 850 consecutive hernias. Ann Surg 238:391–400
Berger D, Bientzle M, Muller A (2002) Postoperative complications after laparoscopic incisional hernia repair. Surg Endosc 16:1720–1723
Bellows CF, Berger DH (2006) Infiltration of suture sites with local anesthesia for management of pain following laparoscopic ventral hernia repairs: a prospective randomized trial. JSLS 10:345–350
Le Blanc KA (2007) Laparoscopic incisional hernia repair: are transfascial sutures necessary? A review of the literature. Surg Endosc 21:508–513
van’t Riet M, de Vos van Steenwijk PJ, Kleinrensink GJ, Steyerberg EW, Bonjer HJ (2002) Tensile strength of mesh fixation methods in laparoscopic incisional hernia repair. Surg Endosc 16:1713–1716
Winslow ER, Diaz S, Desai K, Meininger T, Soper NJ, Klingensmith ME (2004) Laparoscopic incisional hernia repair in a porcine model: what do transfixation sutures add? Surg Endosc 18:529–535
Joels CS, Matthews BD, Kercher KW, Austin C, Norton HJ, Williams TC, Heniford BT (2005) Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture. Surg Endosc 19:780–785
Petter-Puchner AH, Fortelny R, Mittermayr R, Öhlinger W, Redl H (2005) Fibrin sealing versus stapling of hernia meshes in an onlay model in the rat. Hernia 9:322–329
Olmi S, Scaini A, Erba L, Bertolini A, Croce E (2007) Laparoscopic repair of inguinal hernias using an intraperitoneal onlay mesh technique and a Parietex composite mesh fixed with fibrin glue (Tissucol): personal technique and preliminary results. Surg Endosc 21:1961–1964
Lau H (2005) Fibrin sealant versus mechanical stapling for mesh fixation during endoscopic extraperitoneal inguinal hernioplasty. Ann Surg 242:670–675
Conze J, Rosch R, Klinge U, Weiss C, Anurov M, Titkowa S, Oettinger A, Schumpelick V (2004) Polypropylene in the intraabdominal position: influence of pore size and surface area. Hernia 8:365–372
Schug-Paß C, Tamme C, Tannapfel A, Köckerling F (2006) A lightweight polypropylene mesh (TiMesh) for laparoscopic intraperitoneal repair of abdominal wall hernias: comparison of biocompatibilty with the DualMesh in an experimental study using the porcine model. Surg Endosc 20:402–409
Schug-Paß C, Sommerer F, Tannapfel A, Lippert H, Köckerling F (2008) Does the additional application of a polylactide film (SurgiWrap®) to a lightweight mesh (TiMesh light®) reduce adhesions in laparoscopic intraperitoneal implantation? Experimental results obtained in a porcine model. Surg Endosc 22:2433–2439
de Virgilio C, Elbassir M, Hidalgo A, Schaber B, French S, Amin S, Stabile BE (1999) Fibrin glue reduces the severity of intraabdominal adhesions in a rat model. Am J Surg 178:577–580
Toosie K, Gallego K, Stabile BE, Schaber B, French S, de Virgilio C (2000) Fibrin glue reduces intraabdominal adhesions to synthetic mesh in a rat ventral hernia model. Am Surg 66:41–45
Martin-Cartes JA, Morales-Conde S, Suarez-Grau JM, Bustos-Jimenez M, Cadet-Dussort H, Socas-Macias M, Lopez-Bernal F, Alamo-Martinez JM, Tutosaus-Gomez JD, Moralez-Mendez S (2006) Prevention of peritoneal adhesions to intraperitoneal prostheses: an experimental study in pigs. Cir Esp 80:214–219
Saed GM, Kruger M, Diamond MP (2004) Expression of transforming growth factor-beta and extracellular matrix by human peritoneal mesothelial cells and by fibroblasts from normal peritoneum and adhesions: effect of Tisseel. Wound Repair Regen 12:557–564
Klinge U, Conze J, Limberg W, Brücker C, Ottinger AP, Schumpelick V (1996) Pathophysiology of the abdominal wall. Chirurg 67:229–233
Scheidbach H, Tamme C, Tannapfel A, Lippert H, Köckerling F (2004) In vivo studies comparing the biocompatibility of various polypropylene meshes and their handling properties during endoscopic total extraperitoneal (TEP) patchplasty: an experimental study in pigs. Surg Endosc 18:211–220
Scheidbach H, Tannapfel A, Schmidt U, Lippert H, Köckerling F (2004) Influence of titanium coating on the biocompatibility of heavyweight polypropylene mesh: an animal experimental model. Eur Surg Res 36:313–317
Schwab R, Schumacher O, Junge K, Binnebösel M, Klinge U, Becker HP, Schumpelick V (2008) Biomechanical analysis of mesh fixation in TAPP and TEP hernia repair. Surg Endosc 22:731–738
Binnebösel M, Rosch R, Junge K, Flanagan TC, Schwab R, Schumpelick V, Klinge U (2007) Biomechanical analyses of overlap and mesh dislocation in an incisional hernia model in vitro. Surgery 142:365–371
Knook MTT, van Rosmalen AC, Yoder BE, Kleinrensink GJ, Snijders CJ, Looman CWN, van Steesel CJ (2001) Optimal mesh size for endoscopic inguinal hernia repair. Surg Endosc 15:1471–1477
Klinge U, Klosterhalfen B, Birkenhauer V, Junge K, Conze J, Schumpelick V (2002) Impact of polymer pore size on the interface scar formation in a rat model. J Surg Res 103:208–214
Garrard CL, Clemets RH, Nanney L, Davidson JM, Richards WO (1999) Adhesion formation is reduced after laparoscopic surgery. Surg Endosc 13:10–13
Gutt CN, Oniu T, Mehrabi A, Büchler W (2004) Fewer adhesions induced by laparoscopic surgery? Surg Endosc 18:898–906
Duron JJ (2007) Postoperative intraperitoneal adhesion pathophysiology. Colorectal Dis 9(Suppl 2):14–24
Acknowledgments
We thank Mr. Heinlein from GfE-Medizintechnik GmbH for technical support in conducting and analyzing the biomechanical experiments.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Schug-Pass, C., Lippert, H. & Köckerling, F. Fixation of mesh to the peritoneum using a fibrin glue: investigations with a biomechanical model and an experimental laparoscopic porcine model. Surg Endosc 23, 2809–2815 (2009). https://doi.org/10.1007/s00464-009-0509-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-009-0509-0