‘Post-Close’ Femoral Arterial Haemostasis at Endovascular Aneurysm Repair using a Dedicated Large-Bore Vascular Closure Device: A Prospective Real-World Audit

  • Clinical Investigation
  • Arterial Interventions
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

Percutaneous access for endovascular aneurysm repair (EVAR) is commonplace, with its attendant benefits. The combination of continued reduction in device profile and evolution of vascular closure device (VCD) design drives successful and safe percutaneous EVAR. A new such VCD is the MANTA Large-Bore Closure Device designed in two iterations for closure of arterial defects from 10 to 25F. We present a prospective audit of 131 large-bore femoral closures using an ‘all-comers’ approach to device selection.

Materials and methods

One hundred and thirty-one large-bore femoral arterial defects were analysed. Both 14F and 18F MANTA VCDs were deployed in this series as per instructions for use. Primary objectives were technical success, particularly successful deployment, and haemostasis achieved. Failures were denoted as failure to deploy; failure to achieve haemostasis was denoted as active bleeding, haematoma, or pseudoaneurysm formation requiring intervention. Later complications assessed were vessel occlusion/thrombosis or stenosis.

Results

Seventy-six patients (65 males and 11 females, age 75.2 ± 8.7 years) underwent a range of procedures including EVAR (n = 66), TEVAR (n = 2), and reinterventions (n = 8) requiring large-bore percutaneous femoral arterial access in 131 groins. Of these, the 14F MANTA VCD was used in 61 closures (defects ranging from 12 to 18F) and the 18F in 70 closures (defects ranging from 16 to 24F). Deployments achieved successful haemostasis in 120 (91.6%), and failures occurred in 11(8.4%) groins.

Conclusions

This study indicates that a post-close approach using the novel MANTA Large-Bore Closure Device can be undertaken successfully to close a range of large-bore femoral arterial defects at EVAR/TEVAR with an acceptable rate of complications.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
EUR 32.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or Ebook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Lee WA, Brown MP, Nelson PR, Huber TS. Total percutaneous access for endovascular aortic aneurysm repair (“Preclose” technique). J Vasc Surg. 2007;45:1095–101.

    Article  PubMed  Google Scholar 

  2. Gheorghe L, Brouwer J, Mathijssen H, Nijenhuis VJ, Rensing BJ, Swaans MJ, et al. Early outcomes after percutaneous closure of access site in transfemoral transcatheter valve implantation using the novel vascular closure device collagen plug-based MANTA. Am J Cardiol. 2019;124:1265–71.

    Article  PubMed  Google Scholar 

  3. Minion DJ, Davenport DL. Access techniques for EVAR: percutaneous techniques and working with small arteries. Semin Vasc Surg. 2012;25:208–16.

    Article  PubMed  Google Scholar 

  4. Smith TA, Ramirez DE. Awake, percutaneous repair of a ruptured abdominal aortic aneurysm. Ochsner J. 2013;13:248–51.

    PubMed  PubMed Central  Google Scholar 

  5. Lachat ML, Pecoraro F, Mayer D, Guillet C, Glenck M, Rancic Z, et al. Outpatient endovascular aortic aneurysm repair: experience in 100 consecutive patients. Ann Surg. 2013;258:754–8 (Discussion 8–9).

    Article  PubMed  Google Scholar 

  6. Georgiadis GS, Antoniou GA, Papaioakim M, Georgakarakos E, Trellopoulos G, Papanas N, et al. A meta-analysis of outcome after percutaneous endovascular aortic aneurysm repair using different size sheaths or endograft delivery systems. J Endovasc Ther. 2011;18:445–59.

    Article  PubMed  Google Scholar 

  7. Harrison GJ, Thavarajan D, Brennan JA, Vallabhaneni SR, McWilliams RG, Fisher RK. Fascial closure following percutaneous endovascular aneurysm repair. Eur J Vasc Endovasc Surg. 2011;41:346–9.

    Article  CAS  PubMed  Google Scholar 

  8. Chaudhuri A. Femoral arterial haemostasis using an anchored collagen plug after percutaneous EVAR with an ultra-low profile device: prospective audit of an evolving “post-close” technique. Eur J Vasc Endovasc Surg. 2017;54:241–6.

    Article  CAS  PubMed  Google Scholar 

  9. Chaudhuri A. Prospective audit of “post-close” haemostasis following large bore femoral arterial punctures: a second look at the double angio-seal technique. EJVES Short Rep. 2018;40:12–4.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Krajcer Z, Wood DA, Strickman N, Bernardo N, Metzger C, Aziz M, et al. Pivotal clinical study to evaluate the safety and effectiveness of the MANTA vascular closure device during percutaneous EVAR and TEVAR procedures. J Endovasc Ther. 2020;27:414–20.

    Article  PubMed  Google Scholar 

  11. Gupta A, Fouad L, Basir M, Neupane S, Zaidan M, Koenig G, et al. Safety and effectiveness of MANTA vascular closure device after large-bore mechanical circulatory support: real-world experience. Cardiovasc Revasc Med. 2020;21(7):875–8.

    Article  PubMed  Google Scholar 

  12. Tarantini G. MANTA dedicated large-bore vessel closure device. Circ Cardiovasc Interv. 2019;12:e008203.

    Article  PubMed  Google Scholar 

  13. Van Mieghem NM, Latib A, van der Heyden J, van Gils L, Daemen J, Sorzano T, et al. Percutaneous plug-based arteriotomy closure device for large-bore access: a multicenter prospective study. JACC Cardiovasc Interv. 2017;10:613–9.

    Article  PubMed  Google Scholar 

  14. Wood DA, Krajcer Z, Sathananthan J, Strickman N, Metzger C, Fearon W, et al. Pivotal clinical study to evaluate the safety and effectiveness of the MANTA percutaneous vascular closure device. Circ Cardiovasc Interv. 2019;12:e007258.

    Article  PubMed  Google Scholar 

  15. Megaly M, Sedhom R, Abdelmaseeh P, ElBebawy B, Goel SS, Karam J, et al. Complications of the MANTA closure device: insights from MAUDE database. Cardiovasc Revasc Med. 2022;34:75–9.

    Article  PubMed  Google Scholar 

  16. Wong YH, De Backer O, Sondergaard L, Bieliauskas G. Percutaneous management of an embolised MANTA large bore arteriotomy closure device. EuroIntervention. 2019;15:74–5.

    Article  PubMed  Google Scholar 

  17. Hoffmann P, Al-Ani A, von Lueder T, Skoe OW, Tran TT, Opdahl A. Late pseudoaneurysm after access site closure with Manta in transfemoral aortic valve implantation. EJVES Short Rep. 2019;42:34–6.

    Article  PubMed  Google Scholar 

  18. Au SY, Fong KM, Ng WG, Lee KM, So SO, Leung KA. Real-time ultrasound-guided bedside closure of arteriotomy wound using MANTA closure device during venoarterial extracorporeal membrane oxygenation decannulation. Perfusion. 2020;36(2):118–21 (267659120932429).

    Article  PubMed  Google Scholar 

  19. Sarathy K, Khanna V, Baumbach A, Mullen M. Sonographic guidance of vascular closure with MANTA. JACC Cardiovasc Interv. 2019;12:2437–8.

    Article  PubMed  Google Scholar 

  20. Biancari F, Romppanen H, Savontaus M, Siljander A, Makikallio T, Piira OP, et al. MANTA versus ProGlide vascular closure devices in transfemoral transcatheter aortic valve implantation. Int J Cardiol. 2018;263:29–31.

    Article  PubMed  Google Scholar 

  21. Hoffmann P, Al-Ani A, von Lueder T, Hoffmann J, Majak P, Hagen O, et al. Access site complications after transfemoral aortic valve implantation—a comparison of Manta and ProGlide. CVIR Endovasc. 2018;1:20.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Moriyama N, Lindstrom L, Laine M. Propensity-matched comparison of vascular closure devices after transcatheter aortic valve replacement using MANTA versus ProGlide. EuroIntervention. 2019;14:e1558–65.

    Article  PubMed  Google Scholar 

  23. Chaudhuri A. A “post-close” technique for femoral hemostasis after percutaneous EVAR using an ultra-low-profile endoprosthesis system. EJVES Short Rep. 2015;29:18–20.

    Article  Google Scholar 

  24. Ali N, Blackman DJ, Cunnington M, Malkin CJ. Use of the MANTA device to rescue failed pre-closure following transfemoral transcatheter aortic valve implantation. J Cardiol Cases. 2019;19:81–4.

    Article  PubMed  Google Scholar 

  25. van Gils L, De Jaegere PP, Roubin G, Van Mieghem NM. The MANTA vascular closure device: a novel device for large-bore vessel closure. JACC Cardiovasc Interv. 2016;9:1195–6.

    Article  PubMed  Google Scholar 

  26. Kastengren M, Svenarud P, Kallner G, Settergren M, Franco-Cereceda A, Dalen M. Percutaneous Vascular Closure Device in Minimally Invasive Mitral Valve Surgery. Ann Thorac Surg. 2020;110:85–91.

    Article  PubMed  Google Scholar 

  27. Stomp W, Eefting D, van Schaik J, Sudiono DR, van der Meer RW. Antegrade common femoral artery access site closure using the MANTA vascular closure device. Radiol Case Rep. 2020;15:2205–7.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Montero-Cabezas JM, van der Meer RW, van der Kley F, Elzo Kraemer CV, Lopez Matta JE, Schalij MJ, et al. Percutaneous decannulation of femoral venoarterial ECMO cannulas using MANTA vascular closure device. Can J Cardiol. 2019;35(796):e9–11.

    Google Scholar 

  29. Ruck A, Eriksson D, Verouhis D, Saleh N, Linder R, Corbascio M, et al. Percutaneous access and closure using the MANTA vascular closure device in transaxillary transcatheter aortic valve implantation. EuroIntervention. 2020;16:266–8.

    Article  PubMed  Google Scholar 

  30. De Palma R, Ruck A, Settergren M, Saleh N. Percutaneous axillary arteriotomy closure during transcatheter aortic valve replacement using the MANTA device. Catheter Cardiovasc Interv. 2018;92:998–1001.

    Article  PubMed  Google Scholar 

  31. Latsios G, Toutouzas K, Karanasos A, Synetos A, Drakopoulou M, Aggeli C, et al. Trans-femoral TAVI: successful hemostasis of a totally calcified femoral artery (“calcium tube”) with the Manta(c) device. Hellenic J Cardiol. 2020;62(2):158–60.

    Article  PubMed  Google Scholar 

  32. Chaudhuri A. Newer-generation vascular closure devices: plugging the hole in the artery, what about the one in the purse? J Endovasc Ther. 2020;27:977–8.

    Article  PubMed  Google Scholar 

Download references

Funding

This study has not received any funding.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. Chaudhuri.

Ethics declarations

Conflict of interest

The author has received honoraria following educational events from Teleflex.

Ethical Approval

For this type of study, formal consent is not required.

Informed Consent

For this type of study, informed consent is not required.

Consent for Publication

For this type of study, consent for publication is not required (there is no patient-identifying data).

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chaudhuri, A. ‘Post-Close’ Femoral Arterial Haemostasis at Endovascular Aneurysm Repair using a Dedicated Large-Bore Vascular Closure Device: A Prospective Real-World Audit. Cardiovasc Intervent Radiol 46, 835–843 (2023). https://doi.org/10.1007/s00270-023-03437-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-023-03437-7

Keywords

Navigation