Summary
We developed and validated a sensitive and low sample volume liquid chromatographic-mass spectrometric (LC-MS) method for determination of Promethazine hydrochloride in human plasma (0.5 ml) or urine (0.1 ml). The lower limit of quantification in human plasma and urine was 1.00 ng/ml. The inter- and intra-day precisions (CV %) in both plasma and urine were lower than 10%, the mean method accuracies and recoveries from spiked plasma samples at three concentrations were more than 97%. The developed method was successfully applied to determine Promethazine hydrochloride in human plasma and urine, and proved suitable to clinical pharmacokinetic study.
Similar content being viewed by others
References
Tarkkila P., Torn K., Tuominen M., Lindgren L., (1995): J. Acta Anaesthesiol. Scand., 39. 983.
Scherl E.R., Wilson J.F., (1995): J. Headache. 35, 256.
Buzdar A.U., Esparza L., Natale R., Cody R., Calzone K., Benson A.B., Sheehan T., Berry W., (1994): Am. J. Clin. Oncol., 17, 417.
Terndrup T.E., Dire D.J., Madden C.M., Davis H., Cantor R.M., Gavula D.P., (1991): J. Ann Emerg. Med., 20, 31.
Conner J.T., Bellville J.W., Wender R., Wapner S., Dorey F.J., Katz R.L., (1977): J. Anesth Analg., 56, 801.
Ron M., Menashe M., Hochner-Celnikier D., Palti Z., (1984): Eur. J. Obstet. Gynecol. Reprod. Biol., 18, 25.
Lackner J.R., Graybiel A., (1994): J. Clin Pharmacol., 34, 644.
Campbell M., Bateman D.N., (1992): J. Clin Pharmacokinet., 23, 147.
Stavchansky S., Wallace J.E., Geary R., Hecht G., Robb C.A., Wu P., (1987): J. Pharm Sci.. 76, 441.
Ponder G.W., Stewart J.T., (1995): J. Pharm. Biomed. Anal., 13, 1161.
Zaman R., Honigberg I.L., Francisco G.E., Kotzan J.A., Stewart J.T., Brown W.J., Shah V.P., Pelsor R.R., (1986): J. Biopharm Drug Dispos., 7, 281..
Schwinghammer TL., Juhl R.P., Dittert L.W., Melethil S.K., Kroboth F.J., Chung V.S., (1984): J. Biopharm. Drug Dispos., 5, 185.
Fox A.R, McLoughlin D.A., (1993): J. Chromatogr., 631, 255.
Wallace J.E., Shimek E.L, Jr, Harris SC, Stavchansky S., (1991): J. Clin. Chem., 27, 253.
Tompsett S.L., (1968): J. Acta Pharmacol.Toxicol., 26, 298.
Martin E.A., (1966): Can. J. Chem., 44, 1783.
Vanapalli S.R., Kambhampati S.P., Putcha L., Bourne D.W., (2001): J. Chromatogr. Sci., 39, 70.
Guidance for industry, Bioanalytical Method Validation, US Department of Health and Human Services, May 2001.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Liu, P., Liang, S., Wang, BJ. et al. Development and validation of a sensitive LC-MS method for the determination of Promethazine hydrochloride in human plasma and urine. Eur. J. Drug Metabol. Pharmacokinet. 34, 177–184 (2009). https://doi.org/10.1007/BF03191171
Accepted:
Issue Date: