Measurement of heart-rate variability: Part 2-hardware digital device for the assessment of heart-rate variability

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

Part 1 of the paper was dedicated to some general aspects of h.r.v. measurements. In Part 2, the spectral properties of the integral pulse frequency modulator (i.p.f.m.) are shown. The choice of a hardware device for lowpass filtering the cardiac event series by means of a stepwise convolution is discussed. A method to overcome the problems of distortion due to spurious spectral components in the filter passband is shown. The hardware realisation of the filter is described and results are given.

Sommaire

La première partie du texte était consacrée à quelques aspects généraux de la mesure de la v.r.c. Dans cette partie, les auteurs exposent les propriétés spectrales du modulateur intégrateur d'impulsions en fréquence. L'article examine le choix d'un matériel pour le filtrage passe-bas de la série d'événements cardiaques au moyen d'une circonvulation pas-à-pas. Il décrit une méthode pour résoudre les problèmes de distorsion dus aux composants spectraux parasites dans la bande passante du filtre. Il décrit enfin la fabrication du filtre ainsi que de résultats.

Zusammenfassung

Teil I des Textes widmete sich einigen allgemeinen Aspekten von HRV-Messungea. In diesem Teil werden die Spektraleigenschaften des integralen Impulsfrequenzmodulators (IPFM) gezeigt. Behandelt wird außerdem die Wahl einer Vorrichtung zur Tiefpaßfilterung der Herzätig-keitsreihe mit Hilfe einer stufenweisen Windung. Eine Methode zur Überwindung der Probleme der Verzerrung infolge störender Spektralkomponenten im Filterpaßband wird gezgeigt. Eine Hardware-Realisation des Filters wird beschrieben und die Resultate angegeben.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
EUR 32.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or Ebook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price includes VAT (Germany)

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Bayly, E. J. (1968) Spectral analysis of pulse frequency modulation in the nervous system.IEEE Trans. BME-15, 257–265.

    Google Scholar 

  • Coenen, A. J. R. M. (1975) A hardware device for the assessment of fluctuations in heart rate (in Dutch). Ir thesis, Department of Electrical Engineering, Delft University of Technology.

  • Cox, J. R., Fozzard, H., Nolle, F. M. andOliver, G. C. (1969) Some data transformations useful in electrocardiographyin:Stacy &Waxman (Eds.)Computers in biomedical research,3, 181, AC Press, New York.

    Google Scholar 

  • Koenderink, J. J. andvan Doorn, A. J. (1973) Event train decoders with many inputs; pulse density versus momentaneous frequency.Kybernetik 13, 215–222.

    Article  Google Scholar 

  • Lee, Y. W. (1960) Statistical theory of communication. Wiley, New York.

    Google Scholar 

  • Rompelman, O., Kitney, R. I. andCoenen, A. J. R. M. (1975) Measurement of heart rate variability: Part 1.Med. & Biol. Eng. & Comput. 15.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Coenen, A.J.R.M., Rompelman, O. & Kitney, R.I. Measurement of heart-rate variability: Part 2-hardware digital device for the assessment of heart-rate variability. Med. Biol. Eng. Comput. 15, 423–430 (1977). https://doi.org/10.1007/BF02457997

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02457997

Keyword

Navigation